A bench needs two signal generators to share a reference. A coaxial 10 MHz OCXO works with a Tektronix AFG31000 but a Keysight 33500B does not report lock after moving the same source. Both instruments are designed for nominal 10 MHz reference operation. Compare receiver lock range, voltage at the actual input and installed options before replacing the reference.
Compare the reference input specifications and options
Tektronix specifies 1 kΩ input impedance, 100 mVpp to 5 Vpp, and a 10 MHz ±35 kHz lock range for the AFG31000 external reference input. Use the reference connector and the external clock selection. Adjacent connectors can have different impedances and signal functions.
The Keysight 33500B and 33600A data sheet lists an external reference frequency range of 10 MHz ±20 Hz for standard configurations and ±1 Hz for the OCX option. The reference level is 200 mVpp to 5 Vpp. Its input is AC-coupled with a stated impedance of 1 kΩ in parallel with 20 pF. Use the manual and option information for the instrument in front of you.
A hypothetical 10 MHz OCXO warming up at +0.5 ppm is +5 Hz high. That is within the Tektronix and standard Keysight frequency bounds cited above, but beyond the documented ±1 Hz range for the Keysight OCX option. Actual capture also depends on amplitude, waveform, temperature and instrument conditions. At 10 MHz, 0.1 ppm corresponds to 1 Hz. Include the required warm-up interval and the offset during that interval, not just the settled stability, when comparing oscillator candidates.
Check the voltage under the actual load
For a sine reference specified at 0 dBm into 50 Ω, the load dissipates 1 mW. That is approximately 0.224 Vrms or 0.632 Vpp. A source that can be modeled as 50 Ω may give nearly twice its matched-load voltage into a very high impedance. Cable effects and input capacitance still matter. Measure at the receiver and document cable, termination, waveform and probe setting. Other oscillator output structures do not necessarily follow that particular source model.
A measured 150 mVpp at the receiver meets the quoted Tektronix amplitude range and is below the quoted Keysight minimum. Measure under the furthest cable, actual distribution load and supply conditions. A tee feeding another 50 Ω input or a termination can change the level. Confirm that loading is permitted for the source. Use an appropriate distribution stage if the setup needs multiple loads, and keep active source outputs isolated from one another.
Separate frequency from level during troubleshooting. Replace the candidate briefly with a reference that has already locked at that receiver, without changing the cable or route. Keep amplitude within a known interval while comparing frequency. If the offset is acceptable but only the weak signal fails, measure progressively along the distribution chain. Change one variable per comparison.
Use an independent timebase to measure the candidate itself. A counter locked to the same source can conceal common frequency error. To evaluate a 1 Hz range near 10 MHz, verify the measurement uncertainty, including resolution and reference accuracy. Keep a warm-up curve rather than a single settled reading.
Verify lock and phase separately
Record an output measured with the internal reference. Select external reference and retain the lock status and output result. On an isolated bench, an approved change between two known-good reference frequencies can help confirm the output follows the external source. Observe the indicated state when the external reference is removed and after a restart. A passing warm bench test does not document cold start and recovery behavior.
Two instruments locked to the same 10 MHz reference can still require a trigger and alignment procedure for reproducible output phase. Use the manufacturer's supported synchronization method for each instrument and measure at the actual receiving points. A 5 ns mismatch corresponds to 18 degrees on a 10 MHz waveform. Reference distribution, trigger paths and signal cables can each contribute a different part of that delay.
The procurement record should include the tested instrument models and options, allowable reference offset, minimum receiver amplitude, warm-up conditions and load. Engineering can retain the actual receiver trace, lock indication and output comparisons. Verify sustained reference lock for frequency traceability. Where the setup also requires phase alignment, check the trigger and path delay separately and repeat after restart.