Two 10 MHz OCXOs measure apart individually, then lose their beat at the mixer. How do you diagnose it?

Two independent 10 MHz OCXOs differ by a fraction of a hertz when measured separately. Connect them to a mixer for a clearer comparison and the low-frequency output barely moves. The beat may fall below the acquisition path's passband. Alternatively, leakage through the inputs may pull the two sources together until they lock. Keep the mixer ports properly terminated and check whether the two oscillators change frequency upstream of the mixer.

Verify that the bench can see the low beat frequency

Warm both sources and measure each with its rated load and an independent counter timebase. If one of the OCXOs normally serves as the counter reference, document that arrangement. To diagnose unwanted locking, use a third reference outside the coupled paths or a verified isolated observation path. Use buffered outputs where available. An extra instrument tee can change the source load if both branches terminate it.

First drive the mixer from two adequately isolated calibration sources with a known offset. If the bench must measure 0.20 Hz, check the entire chain from the mixer output to the digitizer at that difference. A mixer may translate to low frequency while an output transformer, coupling capacitor or AC input setting rejects a slow beat. Capture several cycles and retain the raw waveform before accepting a displayed zero.

Check the mixer port definitions as well. RF, local oscillator and intermediate-frequency ports may have different drive, termination and frequency requirements. Start with a larger beat and step down using the same wiring. If the measured beat amplitude falls away while independent high-frequency measurements remain separated, investigate the IF port, coupling and acquisition window. The test can identify a deficient bench before either sample is replaced.

Retain the terminations and increase reverse isolation

NIST cautions that leakage between mixer inputs can make the results look misleadingly good. It recommends well-buffered oscillator outputs. Compare the setup after adding buffers with verified reverse isolation. Where needed, adjust attenuation and gain downstream without changing the oscillator load or taking the mixer inputs outside their rated levels. Check the buffers for relevant drift and phase disturbance using a known source before using them for stability measurements.

Compare direct connection, higher isolation and restoration of the direct connection. Observe the two independent high-frequency signals and mixer beat together. If the oscillator has no independent monitor connector, use a directional sample port suited to the circuit. To test one source alone, disconnect the other and place the specified match on the opened mixer or buffer port. After reconnecting it, also try separate supplies and keep cable and shield positions fixed to help identify another coupling path. Look for a repeatable relationship between isolation and measured frequency at the source.

Measure buffer reverse isolation under its specified supply and impedance conditions. Keysight defines it as transmission from output back to input. Test mixer port isolation using the component's intended operating conditions, including where appropriate a comparison with and without local oscillator drive. A standalone buffer result cannot account for paths through the harness or instrument rack. Repeat the test with the complete fixture in place.

Use beat period and isolation as acceptance evidence

Suppose separate measurements show 10 000 000.08 Hz and 9 999 999.88 Hz. The free-running difference is 0.20 Hz, a fractional difference of 2.0 × 10⁻⁸, or 20 ppb. The normal beat period is 5 s, so a 60 s record should contain about 12 cycles. If the connected high-frequency outputs remain bounded in relative phase and the separation returns when isolation increases, the result supports pulling in the comparator. Near a locking boundary the phase may slip intermittently. Extend the recording if a short record appears stationary.

Suppose one reverse path measures 35 dB of isolation with a +7 dBm source. Leakage attributable to that path is approximately −28 dBm. A further measured 25 dB of isolation reduces the estimate to −53 dBm under comparable conditions. These powers refer to the stated connector planes. They cannot establish an internal locking threshold. Impedance and other leakage paths can limit how much the measured result improves. Qualify the fixture near the smallest difference expected and retain drive levels, independently measured frequency separation and beat records.

If one 10 MHz OCXO has an approved electronic tuning input, repeat the connection test at a few known free-running offsets. Record where the measured slope begins to depart from its isolated value and repeat with better isolation. This shows how much margin remains near the smallest frequency difference seen in the project. Freeze ground, tuning and temperature conditions before comparing lots if any of them affect the result. A stronger substitute output may simply reverse the direction of the pulling.

When unwrapping digital beat phase, check for missing samples and log the carrier power in both paths. A weak IF waveform can make crossing time ambiguous. Mark periods with unreliable crossings on the raw record and use the independent RF observation to determine whether relative phase actually stopped. This retains a common comparison record when the mixer or processing changes.