A deep phase-noise notch on a 10 MHz OCXO calls for a splitter check

A single receiver gives a relatively high but smooth phase-noise trace for one 10 MHz OCXO. Connect two receivers and increase the number of cross-spectrum averages. Now a deep notch grows within one band of offset frequencies. Put its bottom into an acceptance report and the sample gains more than ten decibels of apparent margin. Start with the measurement chain. Noise can add constructively between the two receivers, or it can cancel in the cross spectrum.

What survives averaging

Phase fluctuations from the DUT reach both receivers through the power splitter and normally have the same sign after aligning phase. Independent noise added inside the two receiver chains has no such correlation. Its residual statistical scatter falls with averaging. A power splitter and its associated circuits also generate thermal noise. One physical noise source can appear with opposite signs at the two ports. Averaging can reveal that anticorrelated contribution more clearly, and the contribution can cancel part of the source noise. NIST demonstrated the effect with a comparison that varied splitter temperature. A receiving lab can use a comparison with a characterized distribution network.

Consider an idealized example. Suppose the true phase noise of the 10 MHz OCXO is −155 dBc/Hz at one offset. After phase alignment, an anticorrelated term is exactly ninety percent of the source contribution. The real cross spectrum retains ten percent of the positive term. Expressed as a power ratio in decibels, the result is −165 dBc/Hz. Increase the opposing term further and the real part can become negative. An instrument displaying only magnitude can still draw an apparently ordinary trace. The calculation uses spectral densities defined against the same reference. Directly subtracting the two decibel readings does not model the cancellation.

Save the real and imaginary parts, or at least the phase or correlation diagnostics the instrument supplies. Use an identifiable phase-modulation line to confirm the relative sign of the receivers first. A disturbance from the same source should appear with the expected sense on both paths. That makes a later sign reversal more informative. A magnitude-only display still allows fixture comparisons, but those comparisons can implicate the measurement chain without pinpointing a particular component.

Change the splitter and preserve the carrier levels

Record the original setup. Note the carrier level at each receiver, the port assignments, the unused-port condition, receiver gain, offset span, and number of averages. Then install a splitter with characterized isolation and verify that it covers the carrier frequency. Keep adapters and cables at the same reference planes. Check the carrier levels separately and use appropriate compensation if necessary to keep them close. Gain changes can shift the receiver floor or operating region. A large simultaneous change in level makes the notch difficult to interpret.

A reversal test helps. Measure with the first setup, repeat with the alternate one, and then restore the first. If the notch returns with the original fixture, keep all three traces and timestamps. With input conditions satisfied for both receivers, swapping the splitter output lines can help distinguish a receiver issue, one cable, and the whole distribution network. A swap changes some complex spectral features, but cancellation inside the splitter can survive the swap. One port swap alone therefore cannot rule it out.

A single-channel trace gives supporting evidence when taken at an appropriate carrier level and on a suitable receiver path. Its higher noise floor can cover a low-noise notch. If the characterized floor is above the suspect portion of the dual-channel trace, record insufficient visibility. Failing to observe a notch on the single channel does not validate its depth.

Report decisions only at credible offsets

If the OCXO specification calls for readings at one hundred hertz, one kilohertz, and ten kilohertz, assign a measurement status to each point. Where two validated configurations agree, use the measurement uncertainty in the acceptance decision. Where cancellation is suspected, retain the original data and hold the result. Averaging two fixture-dependent readings separated by ten or more decibels only turns the fixture discrepancy into a plausible-looking middle value.

Rule out actual narrowband behavior too. If changing fixtures does not move or reshape the notch, check phase lock settings, carrier level, and calibration status, then sweep a narrower span. If replacing the low-noise reference shifts an anomaly in absolute frequency, investigate the reference path. If the feature recurs solely at one supply, operating mode, or control condition, investigate that operating point on the sample. Each comparison changes one of the fixture, reference, and DUT conditions. Change one at a time so others can reproduce the evidence.

A reviewer needs to know which part of the spectrum is credible and which signs or anomalies the plotting software may have hidden. Retain the splitter identification and assignments, phase-alignment record, carrier levels, number of averages, and the reversal. A quieter measurement floor is meant to reveal the OCXO. A lower trace caused by mutual cancellation cannot establish the sample performance.