A nearby relay energizes and the 10 MHz OCXO moves slightly off frequency. When the relay releases, the reading slowly returns. A relay coil produces a magnetic field and heat, can pull down a supply, and may create a mechanical or electrical transient. Keep the actual relay in one state and use a separate bench source of measurable, reversible static field. Compare the bench response with the field measured at the oscillator in its installed position.
Apply a field while holding the electrical conditions fixed
Mount the oscillator or a representative powered board on a nonmagnetic support. Power a calibrated coil separately. Fix a three-axis probe near the sample position and record field magnitude and direction before and after placing the unit. Immobilize the output coax and supply wires. Keep the reference oscillator and counter away from significant field. Once both the reference and the 10 MHz OCXO are warm, record frequency and case temperature with the coil off to establish natural drift.
Try coil off, positive field, off, negative field and off again. Reverse the order on the next run. Coil off still leaves Earth's field and nearby magnetized structures, so log actual field components for each condition. Set levels within confirmed equipment and component boundaries. Exclude the switching transient from the steady field comparison. At each step record coil current, case temperature, supply at the part and frequency, then use equal windows in the steady portions.
Change field direction by driving other coil axes where possible, with the sample and cables fixed. If you must rotate the oscillator, first repeat the same rotation without an applied field to see how changing the direction of gravity and disturbing the connections affect the measurement. Check the axes relevant to installation. Assembly tolerances can move the device into a worse field direction or position, so retain the full response.
Calculate the field-reversing and common responses separately
NIST describes the typically weak intrinsic magnetic sensitivity of quartz and other possible mechanisms involving magnetic supports, nearby inductors and supply circuits. A specification hosted by Microchip defines magnetic-field sensitivity as fractional frequency change per unit field. A device may show a response that changes sign and one that remains under both polarities. Save the coil-off readings as well as both active states.
Suppose baseline frequency is interpolated from coil-off readings and the measured field increment is +0.30 mT or −0.30 mT along one axis. In this example the 10 MHz OCXO moves +0.010 Hz and −0.002 Hz, respectively. The half-difference is +0.006 Hz. Dividing by 0.30 mT gives a finite-range slope of +0.020 Hz/mT, or a fractional sensitivity of +2.0 × 10⁻⁹/mT. The half-sum is +0.004 Hz, a fractional shift of +4.0 × 10⁻¹⁰. The first quantity captures the polarity-reversing response. The second includes effects present in both conditions, potentially including heating.
If the installed field can rise by +0.30 mT and the added-frequency allowance is ±0.008 Hz, the measured +0.010 Hz fails. The half-difference alone would pass. The slope describes the tested range. Magnetic remanence or nonlinear response can make extrapolation unreliable. Check every return to coil-off for residual frequency error and report repeated residuals with return timing and dwell conditions.
Control coil heating and the real relay supply independently
Equal positive and negative current magnitudes give similar coil dissipation and help identify the sign-changing contribution. The common offset needs another control. Use an approved low-field condition with similar heating or back the coil away, while monitoring temperature near the oscillator. An immediate frequency reversal before appreciable temperature change supports a static field effect. A slower same-direction drift after the case warms keeps heating in the diagnosis.
Return to the production relay and measure all three field axes at the oscillator, together with supply transients, steady voltage drop and temperature. If an externally applied field reproduces the installed frequency shift and removal restores the original reading, compare relocation, orientation and an approved shielding option. A substitution report should capture measured field, orientation, dwell, maximum shifts at both polarities and repeatability. If the power path produces a similar shift, verify electrical isolation before concluding that the 10 MHz OCXO needs replacement.
If the test bench controls frequency through the tuning pin, put that loop in an approved test mode for the entire sequence. You can hold the tuning voltage at a documented value and monitor the actual pin independently. If the system must remain closed-loop, save its control voltage beside the case temperature because correction can hide some of the frequency shift. If oven telemetry is available, check whether it shifts as the field is applied before separating the OCXO from its support electronics. Confirm all test-point definitions and loading limits first.
Run a field-off and field-on check on the reference chain. Keep a reference source outside the field and send its signal through a suitably matched substitution path near the coil. If the comparison still jumps when coil drive changes, inspect probe routing, grounding and instrument triggering. A millihertz-level test requires a background run that establishes how stable the setup is without a sample response. Report effects below measurement resolution separately from clearly over-budget shifts so another shift can reproduce the decision.