A comparator shows that the frequency difference between two 10 MHz OCXOs moves over time. Allan deviation describes the pair, but it does not assign a contribution to each source. With a third free-running reference of comparable stability, measure three pairs simultaneously and apply the three-cornered hat. It requires little correlation on the averaging times of interest and a measurement system whose own noise is adequately low.
Measure all three pairs during the same interval
Warm A, B and C. Acquire A versus B, A versus C and B versus C with a validated parallel phase comparison. The NIST frequency stability handbook calls for simultaneous records. Three measurements made on different evenings experience different equipment and environmental conditions. Subtracting their variances can give an invalid result. Use common timestamps and mark gaps or instrument resets.
Keep the three 10 MHz OCXOs independent. If they are all steered by one reference or share a strong thermal disturbance, they can move together. Subtraction between pairs removes some of that motion and can underestimate individual instability. Check supplies, distribution and reverse isolation before measuring so the oscillators do not pull each other through the comparator. Record temperature, voltage and any external steering.
Before the run, split one source into the measurement paths to check their background and relative delay. If all three differences are measured independently, align signs and timestamps and check whether their sum is close to zero. If the third difference is derived from the first two, closure is automatic and says nothing about the comparator. Measurement noise from each path can still enter the result.
Calculate the three pair variances at each averaging time
Use the same sample interval and Allan deviation estimator for all pairs. For example, evaluate the required results at one, ten and one hundred seconds. Use matching valid intervals. Allan deviation is the square root of variance, so square the deviations before adding or subtracting them.
Suppose the pair deviations at ten seconds are 5, 6 and 7 in units of 10⁻¹¹, for AB, AC and BC respectively. With uncorrelated sources, the A variance is half the sum of AB and AC variances minus BC variance. This gives (25 + 36 − 49) ÷ 2 = 6 in units of 10⁻²². Taking square roots gives estimated deviations of approximately 2.45 × 10⁻¹¹ for A, 4.36 × 10⁻¹¹ for B and 5.48 × 10⁻¹¹ for C.
If the project limit at ten seconds is 3.5 × 10⁻¹¹, A is a candidate for confirmation. Check confidence, instrument background and the other required averaging intervals before approving parts. If the procedure requires a separate low-noise reference, run that confirmation and retain the three paired records for diagnosis.
Treat a negative variance as a failed estimate
If the third deviation is 8 rather than 7 and the others are unchanged, A gives (25 + 36 − 64) ÷ 2 = −1.5 in units of 10⁻²². A negative variance provides no reportable Allan deviation. Setting it to zero would imply unsupported performance. NIST identifies strongly different stabilities, correlation and insufficient data as possible causes, especially at long averaging times. Check simultaneity, shared environmental effects, reference coupling and instrument noise.
Retain raw and processed series for long averaging times. If linear drift was removed for diagnosis, apply the same method to each pair and keep both results. Exchange comparison channels to see whether the estimated instability follows the instrument. A variance that turns negative only after one oscillator moves to another input warrants checks of that path and its sample coverage.
Record how the three 10 MHz OCXOs were powered and connected, the valid sampling intervals, estimator and usable sample count. Flag negative or borderline estimates as inconclusive and use a longer synchronized run or quieter reference. The three-cornered hat can guide further verification and expose limitations of the test setup.
Compare positions and instrument paths separately. Keep oscillator positions fixed and rotate comparator channels to see whether the ranking follows a connector. Then keep channel assignments fixed and exchange enclosure positions where permitted. This checks instrument paths and airflow separately. If ambient conditions change across the test positions, do not combine them as one operating condition.
A logger may drop a frame and still pair samples by row number. Check timestamps and verify the spacing required by the estimator. Recompute using one contiguous interval that all channels share where possible. Filling missing points separately with normal-looking values can bias the combined variance.
Record actual observation span for each averaging time. Adequate data at ten seconds does not give equivalent confidence at one hundred or one thousand seconds. Include usable samples or effective degrees of freedom. For prioritizing a later calibrated test, use an interval where the estimates resolve a difference. Final acceptance still needs the product's specified averaging times.