How can you tell whether a 125 MHz LVDS oscillator spur was counted twice in the jitter report?

Two instruments measure the same 125 MHz LVDS oscillator. Their noise floors look similar but their integrated jitter differs by several fold. Place the marker on the line 25 kHz from the carrier. One report gives dBc and the other gives dBc/Hz. Check whether the instrument counts the line as separate power, removes it, or integrates it as density. Change the processing with the same device and calibrate the chain with controlled phase modulation to see where the discrepancy originates.

Save the line and noise-floor measurements separately

Keysight's Infiniium phase-noise guide describes three spur treatments. Normalized mode treats trace points as noise density. Omit removes identified lines from the noise integration. Power expresses identified lines in dBc and combines their phase jitter with the continuous noise contribution. Detection depends on the instrument threshold and moving average. Save the instrument version, option, integration band, detection settings, and line table. Check the actual behavior on other instruments, even when their labels look similar.

Connect the 125 MHz LVDS oscillator with the specified differential interface and load. If the instrument only has a single-ended input, use a verified differential adapter and check the carrier level and input range. Record the adapter and observation point and confirm that the chain is not overloaded. Export the trace and spur table together. A trace with the identified spur already removed cannot be used to compare peak heights with one that retains it.

Inject a known phase-modulation line

Consider a 125 MHz carrier with a flat −120 dBc/Hz phase-noise floor from 1 to 101 kHz offset and a single-sideband pure phase-modulation line of −60 dBc at 25 kHz. The continuous single-sideband power ratio is 10⁻¹² × 100,000 = 10⁻⁷. The line power ratio is 10⁻⁶. In the small-angle approximation, double each single-sideband ratio and take its square root to get phase RMS. Divide by 2π × 125 MHz. The noise gives about 0.569 ps and the line gives 1.801 ps. Adding the variances gives a total of about 1.889 ps.

Use a known phase-modulated source as a substitute. Choose a line well inside the band and strong enough for detection, then capture the background with modulation disabled. An RF power spectrum shows the sideband magnitude for amplitude modulation as well. A line of unknown origin on an RF spectrum cannot be directly substituted into the phase-jitter example. Otherwise amplitude modulation coupled from the supply can be reported as timing degradation.

Run Omit and Power with the same integration band, bandwidth, and input level. If identification is consistent, one should approximate the noise-only result and the other should include the added variance. Alternatively, subtract the line variance from the total in Power mode and compare with Omit. If the instrument total already includes the line, adding it again raises the example to about 2.609 ps. Retain that value as a spreadsheet cross-check.

Use two resolution settings to check the line marker

Where the phase-noise or spectrum measurement allows resolution changes, select two bandwidths that still resolve the line from its neighbors. The integrated power of an isolated stable line should agree when the full line is captured. A noise marker that normalizes the peak to one hertz instead can change by about 10 dB for a tenfold change in equivalent noise bandwidth. Window shape, energy spread across bins, and carrier drift affect the check. Measure a power span wide enough to capture the line before drawing conclusions about the oscillator.

Return the bench to its original state before qualifying the real samples. Turn off the reference modulation, restore the defined load, and repeat the measurement over the specified observation time. If a spike appears in only one report, check the detection threshold. Where practical, control the neighboring signal or supply separately and observe whether the spur or only its processing changes. Many applications impose a wideband RMS limit and a separate limit for a single line at a specified offset.

Reduce the calibration source’s modulation and see where its line disappears from the automatic spur table. Consult the instrument’s processing description to interpret what happens to the floor and total when the threshold is changed. Record cases where a power marker still measures the line but the table no longer classifies it separately. Do not combine lot records exported with different thresholds.

When changing the connection, include the carrier level, instrument floor, and repeatability in the comparison. Some instruments also remove predicted internal system spurs. A real DUT line can overlap a removed point. Move the calibration line around that offset to verify the feature and export. If that is an acceptance offset, record the internal-omission setting and keep the DUT and calibration data separately.

Keep the floor integration, discrete lines, and total with spurs in the same record and identify the acceptance metric. If the 125 MHz LVDS oscillator swap leaves the measured floor, line power, and carrier level equivalent, align the counting rules in the instrument and spreadsheet. Once the processing agrees, compare line power under the actual supply and termination if the difference consistently follows the device. The common setup then supports a meaningful oscillator comparison.