A 156.25 MHz oscillator's TIE drops when the reference changes. Can the old report still be used?

One 156.25 MHz active oscillator capture shows a ramp in time interval error. Reprocessing the same waveform produces a small trace around zero. The hardware did not change. The measurement changed the ideal edge assigned to each observed edge. A report with extrema alone cannot establish whether the result describes accumulated error or short-term variation after removal of average frequency offset.

Define the ideal edges

With a fixed nominal reference, specify an origin and timebase. The ideal edges occur one nominal period apart. Number observed edges of the same polarity and subtract ideal arrival times. If the oscillator is fast, observed edges arrive increasingly early and the TIE has a negative slope. A reference based on the captured average frequency or a straight-line fit instead adapts to the record. The fitted period absorbs most of the static frequency difference, leaving deviations from the fitted edge sequence.

In a Tektronix scope example, clock TIE is calculated relative to the average frequency in the capture. Another Tektronix note explains how a recovered clock follows some low-frequency phase motion. Fixed nominal, record-average, and tracking references all have uses. An acceptance report should identify the reference and any recovery settings so named results can be compared.

Suppose a 156.25 MHz active oscillator is fast by 2 ppm, or about 312.5 Hz. The nominal period is 6.4 ns. With a fixed nominal reference, ten milliseconds gives about 20 ns of accumulated advance. That exceeds three periods, but continuous edge indexing still supports the comparison. A phase display that wraps once per cycle can turn the ramp into apparent reversals. Verify numbering and phase unwrapping before reporting a separate frequency offset.

Process the same timestamps twice

Add a sinusoidal timing variation of about 80 ps peak to the example. One full cycle spans about 160 ps peak to peak. A fixed-reference ramp of twenty nanoseconds dominates that variation. After removing a best-fit slope, the small oscillation becomes easier to see. Over a finite record the fit can also absorb some slow motion. A short portion of a sinusoid can resemble a ramp and disappear largely into the fit. Set record length for the lowest component of interest and repeat with longer data.

Check trigger and edge validity in the raw waveform. An extra crossing from reflection, a genuinely missing edge, or a lost acquisition can shift the index by one. Fitting that sequence can redistribute the timing error and create an apparent slow drift on either side. Check raw waveforms around jumps and verify timestamp continuity. Handle any real gaps under the test definition and retain the original record.

To emulate the downstream receiver, reprocess with its applicable clock recovery settings. Bandwidth, loop structure, and transition conditions can change the reference. A wider loop can follow more low-frequency timing motion and remove it in the subtraction. Select the applicable system or interface settings and preserve any comparisons among agreed conditions. Choosing the smallest number weakens comparability.

Verify that two benches mean the same thing

Share raw timestamps, reference timebase, and calculation origin between labs. Use a test source of known offset to confirm TIE sign and slope. A test with known slow timing variation checks how fitting or tracking behaves. Verify calibration and external reference connections. Automatic fallback from a missing external reference to an internal timebase can produce a plausible but different result. Qualify both setups before running the original and substitute.

Keep separate evidence for short-term variation and accumulated time error when comparing 156.25 MHz active oscillators. A substitute with a smaller fitted residual can still have a larger static offset. A synchronization application needs the fixed-reference record and mean frequency. Where downstream tracking is allowed, report the specified recovery settings and receiver decision. Multiple calculations from one capture explain differences between reports without changing the device.

If the old purchase sheet has only a TIE field, append reference type, duration, edge polarity, and data exclusions. Reproduce the older method first and then report the application method. If the agreement specifies both, apply each corresponding limit. A retest on another instrument can then distinguish connection effects from differences in processing.

Alignment at the start determines what is removed. Subtracting the first point removes a fixed phase difference while preserving accumulated slope. That fits a system allowed to align arbitrarily at power-up. If absolute arrival relative to a sync pulse is specified, the initial offset belongs in the result. Preserve sync markers and edge indices across retriggers so independently aligning records cannot hide a switching step. Document whether the averaging interval excludes warm-up or a handover transient.

References