A speaker plays a fixed tone inside a product. Two narrow lines appear around its 20 MHz temperature-compensated oscillator carrier and disappear when the sound stops. The powered speaker can create airborne pressure, structural vibration, and electrical disturbance. All can put energy at the same offset. Before using the result to choose a substitute, measure each possible path and change one in a controlled comparison.
Record pressure, vibration, and power separately
NIST identifies acoustic, vibration, and electrical interference among possible problems in frequency measurements. A Vectron study uses controlled periodic stimuli to measure acceleration and pressure sensitivity in quartz resonators. Its pressure experiment changes air pressure in a chamber and does not directly specify product acoustics. For an installed speaker, apply the idea of controlled stimulus and record pressure near the oscillator, local acceleration, supply, and phase response.
Measure pressure near the oscillator with a calibrated microphone and record position, orientation, and enclosure. Volume settings cannot establish local pressure. Closing a cover or moving the board changes the acoustic field. Place an accelerometer near the device when practical. A reading on the speaker enclosure or baseplate can miss local board motion. Check that the sensor mass and mounting do not materially change the result.
Estimate the scale
Suppose a stable 500 Hz tone gives 80 dB SPL near the oscillator, relative to 20 micropascals RMS. Multiplying the reference by ten thousand gives 0.20 pascal RMS. If phase demodulation gives 0.080 Hz RMS frequency deviation at the same tone, the fractional deviation at 20 MHz is 4 ppb. Dividing by pressure gives about 20 ppb per pascal for this assembly. This includes contributions through the enclosure, board, or measurement chain until those paths are controlled.
For a single 500 Hz frequency modulation, divide RMS fractional frequency deviation by twice pi times modulation frequency to get about 1.27 ps RMS periodic time deviation. Verify phase detector bandwidth and background at that level. A slow counter averaging many cycles may show very little. Retain a phase record with adequate sampling if acceptance concerns periodic sidebands.
Change one path and observe the other channels
Use a separately powered sound source mechanically detached from the board where practical. Alter position or direction and remeasure pressure and acceleration. If both move, the phase change is still ambiguous. For a second control, change the support and restore the same measured pressure, then compare local acceleration and sideband response. Document support and frequency. A change in mounting can introduce its own resonance.
Create an electrical control as well. If the speaker amplifier supports an appropriate silent load, compare that state, the product speaker, and independently produced sound. Record the component on the oscillator supply at the tone frequency. A spur that survives silence calls for checking supply, ground, and instrument coupling. If an acoustic response survives after relevant electrical and mechanical controls, examine routes from the enclosure into the device.
Substitute a stable known source at the same frequency without changing the measurement termination, then repeat the acoustic stimulus. A coax touching the housing or a moving probe stand can affect measured phase. Retain the control with the product record. Matching the offset to the tone indicates association. Changes under a repeatable stimulus provide evidence for the route.
Compare the two parts in the same assembly
Test both 20 MHz TCXO lots on the same board with matching mounting, power, and output conditions at the tones of interest. Report microphone and acceleration locations, support, and phase demodulation. A per-pascal result applies to its stated tone and assembly. Verify approximate linearity with another allowed level. If the response is near instrument background, repeat controls or state the detection limit before ranking parts.
If a substitute fails only with a particular cover installed, qualify it in the final enclosure. A supplier can repeat a baseline and controlled stimulus, while per-pascal results from different fixtures need matching conditions before comparison. A useful acceptance record specifies the setup, allowable spur or time deviation, and paired results.
Repeat at two nearby tones and restore the same pressure next to the device at each. If the response rises in a narrow band while local acceleration rises along one axis, check mounting and structural response. If restoring pressure changes the direction of motion, one acceleration axis may be insufficient. A fixed spur present with no stimulus that stays in place when the tone changes belongs in the bench baseline. Retain evidence for both the tone-dependent line and any local structural resonance.