Multiplying crystal voltage and current gives a negative average on the scope. The 24 MHz passive crystal is operating normally and has no separate source of power. Check current direction, channel delay, and measurement loading. At high frequency voltage and current can be far apart in phase. Dissipation can be a small fraction of apparent power, so a timing error much shorter than one cycle can distort the result.
Measure both terminals and define current direction
Take voltage across the crystal and define positive current entering the positive-voltage terminal. Identify the actual nodes and signs, including the sense element and differential probe polarity. Multiplying a crystal node's voltage to ground by crystal current does not give terminal power. The node voltage can include circuit common mode and bias. Recheck polarity when changing probes or reversing current pickup.
If using a shunt, put it in the branch that carries only the crystal current. A protection path or load capacitor sharing the shunt contributes to the result. Record shunt resistance, parasitics, frequency response, and oscillator frequency after connection. If the sensor alters a high-impedance node, repeat with a qualified lower-burden arrangement. A stable reading alone does not establish an undisturbed circuit.
One nanosecond can flip the sign
Suppose voltage is 0.40 V RMS, crystal current is 0.60 mA RMS, and the phase angle is 84 degrees. Their product is about 240 microvolt-amperes. Multiplying by the cosine gives real power of about 25.1 microwatts. A 24 MHz period is about 41.67 ns. A one-nanosecond channel error can add 8.64 degrees. In that direction the apparent angle becomes 92.64 degrees and reported power is about negative 11.1 microwatts. Opposite timing error raises the result.
This ideal example uses one dominant frequency. It shows why timing near quadrature is sensitive. Without phase calibration, peak-to-peak voltage and current cannot establish dissipation. Align both waveforms, multiply corresponding samples, and average over complete cycles. Check convergence with more cycles. With substantial harmonics, verify channel magnitude and phase response over the relevant band. A single fundamental phase angle does not describe the whole spectrum.
Verify alignment with a known load
Tektronix recommends matching voltage and current probe propagation delays before computing sample-by-sample power. Use the probe manufacturer's deskew fixture or equivalent specified method to present the same event to both paths. Then check at the relevant frequency with a known predominantly resistive load. Cable delay is only part of the problem. The sense element or a filter can add phase shift at 24 MHz. Include the reference-load phase if its reactance is appreciable.
Check offset with no excitation. Follow probe instructions for demagnetization and zeroing, and check voltage-probe common-mode residue. Record bandwidth, sample rate, and digital filtering for both channels. For exported processing keep sample times and interpolation methods. Estimate sensitivity to residual skew for near-quadrature results before using them at a decision limit.
Cross-check against the vendor drive-level method
ST's oscillator note estimates drive level using RMS crystal current and the corresponding resistance at load resonance. Averaged terminal voltage times current can provide a cross-check when the signals represent the same port and phase has been calibrated. A large disagreement calls for checking resistance conditions, current path, phase, and loading. Impedance magnitude cannot replace the specified resistance. A sign error is not evidence that the passive crystal is delivering power.
Compare both 24 MHz crystals under the same supply, temperature, and load. Place the vendor's drive limits at the front of the report. Check the highest applicable drive and verify startup if both parts approach their limit. Agreement with the specified method and a verified reference load gives evidence for a meaningful decision.
Save traces before and after alignment together with probe identities, connection, and correction. A change of probe can restore the error if the old correction is reused. A second technician should be able to reproduce the calculation change without changing the crystal.
Budget residual channel delay. In the example, a hundred picoseconds is about 0.86 degree and changes average power by roughly 3.6 microwatts. If a sample is only two or three microwatts from its limit, this contribution can span it. Check repeatability and the high-frequency standard and include residual skew in the uncertainty record. If incoming acceptance specifies another traceable method, retain the terminal-power experiment as diagnostic evidence and apply the specified method for release.