One of two clock outputs is disabled, yet its counter channel keeps showing 25 MHz. The display could retain the last valid answer, or interference from the other clock could be triggering new readings. Verify that the oscillator stopped at the actual receiver. Check whether the instrument produced a new valid measurement. Then change only the neighboring clock or measurement connection and determine which signal the reading follows.
Crosstalk can move crossings before it creates false counts
The Keysight 53200A time measurement specifications include crosstalk from an adjacent standard input. The typical value applies to its documented conditions. It does not establish isolation across all frequencies, cables, or boards. In an illustrative voltage calculation, −60 dB coupling from a 2 V peak-to-peak aggressor produces 2 mV peak to peak. That may be too small to trigger alone, yet it can move crossings of a weak clock.
A 25 MHz sine at 250 mV peak to peak has a midpoint slope of about 19.6 mV/ns. A 1 mV peak disturbance at the crossing gives a timing movement on the order of 51 ps. Nearly synchronous clocks can make it coherent. A small frequency difference can make the crossing error vary periodically. This is a local slope estimate. Actual counts depend on phase, waveform, and qualification thresholds.
Exercise the two paths in turn
With both running, record selected channel, terminations, ranges, coupling, thresholds, frequencies, and valid measurement state. Disable only the DUT clock while preserving loading. Observe its source and receiver independently. Check whether a retained display actually represents a new capture. If new valid readings continue, move the neighbor from 25 to a verified 24 MHz where supported. If the apparent DUT reading follows, the neighbor reaches the measurement chain somehow.
Keep board operation fixed while removing the neighbor from the other counter channel or applying a suitable termination under the bench procedure. If the reading disappears, investigate counter isolation and cables. If the board's receiver still shows the neighboring frequency, examine routing, returns, distribution, and connectors. Swap channels with equivalent cables and terminations to determine whether the discrepancy follows the bench or board. Avoid combining incompatible active outputs.
Keep enabled and disabled requirements separate
For disabled operation, check the receiver's valid edge threshold and qualifying pulse width. A false counter reading does not automatically establish an actual receiver edge. If the receiver sees real coupled pulses, changing only the counter threshold can conceal them without repairing the board. Archive enable state, edge definition, independent waveforms, termination, and instrument validity.
A replacement with different swing can give a different trigger error on the same bench. Compare with common loading, definitions, and adequate resolution. Restore both clocks and compare single-channel with simultaneous operation. If the artifact occurs only in the bench's dual-input configuration, repeat in a verified setup and preserve the original evidence with its conditions.
Make a negative control with the DUT disabled and the neighbor in normal operation. Keep observation time and detection coverage the same after removing the neighbor. If a correctly terminated bench still reports occasional frequencies with no applied signal, verify input noise and configuration with a known source. The control can reproduce a false report without relying on the final display alone.
After changing instruments, check trigger coupling and display updates. One counter may report a frequency in autorange, while another may hold its last reading after losing an input. Apply a known signal to the same input, timestamp an acquisition, then remove it and observe how long the display or remote status takes to exit its valid state. Allow the full timeout on each switch and save the status or error flag. This establishes whether the procedure can distinguish fresh, retained, and unavailable results. It does not estimate crosstalk on the board.
A production fixture can classify failures so a second technician can reproduce them. A fresh result with both channels loaded and the DUT correctly disabled calls for an independent search for receiver edges. If the fault disappears only when the victim is moved to a spare input with the neighbor still active, check the original input's threshold and the instrument's isolation conditions. Pulses that meet the receiver's frequency and width criteria remain a board timing fault. Qualifying the fixture before comparing substitute oscillator lots keeps one bench defect from appearing as several component defects.