Forward and reverse sweeps give different resonances for a 27 MHz crystal. How long should the VNA dwell?

A fast low-to-high sweep puts the notch of a 27 MHz crystal on one side. Reverse the point order and the notch appears on the other side, with the same fixture and sample. The device may still be responding to the previous stimulus when the analyzer takes data at its new frequency. Compare both directions, several dwell times and settled single-frequency readings before attributing the difference to physical hysteresis.

Hold drive and measurement definition constant

Keysight's network analyzer manual describes distortion when sweep or dwell is shorter than the response of a narrowband device. For stepped measurements, dwell is the wait at each frequency before acquisition. Depending on the instrument, sweep-start delay, trigger delay and IF bandwidth are separate controls. An extra 50 ms before the sweep cannot replace a per-point wait.

Decide whether the target is series resonance at the calibrated plane or loaded resonance under a specified capacitance. Verify the fixture and keep source level, crystal drive and temperature fixed. Do not let changing sweep time also change point count or drive unnoticed. Drive can change the actual crystal response. IF bandwidth changes the instrument's response. Adjust one identified setting at a time.

Save the fast sweep's raw complex data. If direct reverse sweep is unavailable, use a reversed frequency list or step down manually. Check the instrument with a broadband standard if the sweep mode changes. Start repeated passes from a defined point outside the measurement region to control history from the previous stimulus.

Verify convergence at a few representative frequencies

Select several frequencies near the feature and approach each from below and above. Mark each frequency change and compare short, longer and extended waits until complex readings agree within the required tolerance. Then repeat a sufficiently dense sweep using the validated timing. If a single-frequency reading drifts after long waits, check temperature, incident drive and fixture contact before assuming a first-order transient.

For scale, suppose the effective Q of one resonance at 27 MHz is about 90,000. The linear resonator amplitude time constant is approximately Q divided by pi and frequency, which gives 1.06 ms. After 0.20 ms a first-order transient retains around 83% of its starting deviation. After five time constants, around 5.3 ms, the remainder is approximately 0.67%. This single-mode estimate helps choose test durations. It does not account for IF filtering or source settling.

Suppose a continuous sweep covers 3 kHz in 50 ms, or 60 kHz per second. The stimulus then moves about 64 Hz within a 1.06 ms time constant. Extending the sweep to 500 ms reduces that motion to about 6.4 Hz. This shows the scale of a possible fast-sweep problem. It is not a correction for the displayed notch frequency, which depends on dynamic phase and receiver filtering. Use dwell and direction checks to verify the actual response.

Qualify the production timing against repeatability

Suppose the bench requires repeated frequency measurements to agree within 10 Hz. A short sweep gives a 56 Hz direction difference. Longer dwell reduces it to 7 Hz, and another increase gives 3 Hz. The first setting cannot support acceptance. Account for reference and fixture uncertainty before using the slower settings to classify parts. Averaging two distorted notch frequencies assumes errors that may not cancel.

For production throughput, locate the resonance with a rough scan, then dwell longer on a smaller set of points around it. A narrow scan about the principal mode cannot substitute for a spurious-mode check elsewhere. After a change to fixture, IF bandwidth or drive, rerun the slow baseline on a typical sample and a narrow borderline sample. Confirm that the production configuration makes the same decisions.

Archive direction, frequency list, dwell, drive, reference plane and fitted quantity. If a settled bidirectional measurement agrees, keep the fast trace as evidence of the fixture method issue. If a second calibrated instrument still shows a direction effect, investigate actual temperature, drive sensitivity or other physical changes in the sample. A supplier can then reproduce the defined procedure without having to interpret a fast-sweep screenshot.

At one fixed frequency, record the receiver's response over time without continuing the sweep. If the instrument's complex reading still moves, repeat the timing on a calibration path without the sample. The receiver and fixture have their own settling behaviour. A similar blank response directs attention to tuning, automatic level control and IF filtering. Verify the blank first, then repeat with the crystal.

Compare more than the automatic marker. When the resonance lies between measurement points, marker interpolation, smoothing or resampling can change an apparent minimum. Sort both sets of raw complex data by actual frequency and apply the same fit or phase rule. Retain amplitude and phase residuals as well as frequency results. This helps distinguish a marker calculation from distortion already present in acquisition.

References