Two 40 MHz passive crystals start at very different times in the same cold board. In some captures the first signal rises above noise later. In others the rise from small amplitude to a usable signal is slower. Comparing envelope slope in the same small-signal amplitude interval separates changing initial conditions from changing growth rate. Each points to a different board investigation.
Fit growth only in the small-signal interval
Analog Devices describes startup as initial disturbance, small-signal envelope expansion, and final amplitude limiting. With net negative resistance, a linear approximation predicts exponential growth. Plotting the natural logarithm of envelope amplitude against time gives an approximately straight portion. Formal startup time still uses the product's amplitude, frequency, or logic criteria. The slope helps locate why it is slow.
Suppose the first sample takes 80 microseconds to rise from 10 to 40 millivolts in its verified small-signal region. Dividing by the natural logarithm of four gives a time constant of about 57.7 microseconds. A second sample taking 120 microseconds gives about 86.6 microseconds. The second envelope grows more slowly. The absolute time of the first threshold crossing also depends on supply ramp, initial disturbance, and noise.
At one working point, a simplified model gives an envelope time constant of twice motional inductance divided by net negative resistance magnitude. If both crystals have about 8 mH motional inductance, the examples imply net magnitudes of about 277 and 185 ohms. With different inductances, equal slopes can still imply different resistance. These estimates guide the margin investigation. Actual loading, amplifier state, and parasitics vary, so the fit cannot replace the specified margin test.
Verify that the envelope represents oscillation
Use a method supported by the controller or a probe verified to have acceptably low burden at the crystal. Extract amplitude from original samples with bandwidth appropriate for 40 MHz. A power transient can produce a short wideband spike. Verify continuous oscillation at the target after it. A lone amplitude threshold can flag the supply event. Document filter delay and acquisition settings.
Measure bench noise and response using settled operation. Choose a fitting region above noise and below clipping or gain transitions. Compare slopes over two overlapping ranges. If window choice changes the result substantially, check supply ramp, automatic gain, and other frequency components. Avoid compressing an inconsistent whole waveform into one time constant. Retain the envelope and reasons for excluded records.
Separate different effects with repeated cold starts
Hold supply, enable, and specimen temperature fixed. Repeat after the specified off interval. If the first low-amplitude crossing varies while fitted growth stays similar, investigate initial disturbance and reset timing. If a substitute repeatedly grows more slowly through the same region, compare motional parameters, drive, and loading. Swap original and substitute between boards to see whether the effect follows the crystal or assembly.
If the original grows slowly at low supply but behaves normally at room conditions, record the actual pin voltage. The supply display is insufficient. Changing gain or load capacitance can improve growth while altering steady frequency or drive. Verify those results as well. Microchip's oscillator description treats startup gain, equivalent parameters, and small-signal time constant as distinct engineering quantities.
Keep diagnosis with acceptance evidence
For both 40 MHz crystals, retain the cold envelope, fit window, and time of valid clock under a common setup. Apply the product's startup and frequency limits. If envelopes differ but both provide valid clocks on time, check margin across supply, temperature, and samples before specifying a separate slope requirement. If the available captures do not explain the variation, repeat with retained parts and verify that the sensing arrangement itself has not changed startup.
For a lot comparison, align copies of the envelopes at the first crossing of the same low amplitude. This highlights shape while removing the original enable-to-crossing interval. Keep the original timelines as well. If only settled cycles were captured and the early interval is missing, mark the fit unavailable. Visual distance to full scale cannot reconstruct it.
Retain trials with no startup or only brief oscillation. They may lack a valid fitting region and still provide the clearest startup evidence. Record reasons separately for valid fits, failed starts, and incomplete acquisitions. Keeping only successful fits can hide a failure-rate difference despite similar slopes. Include supply checks, measurement zeros, and valid trial counts so missing captures are not mistaken for device failures.
Link each cold-start capture to its retained sample identifier.