A waveform board uses a 50 MHz active oscillator as the AD9834 master clock. On entering a low-power state, its analog output holds an intermediate value. The waveform returns later but its phase differs from a source that never stopped. Check the AD9834 mode before blaming the oscillator. Internal phase accumulation and the output DAC can be disabled independently, with different waveform, current, and phase behavior.
The external clock can run while phase accumulation stops
The Analog Devices data sheet distinguishes two power-down controls. SLEEP1 stops the internal clock and phase accumulation. With the DAC still enabled, the analog output stays at its current value. SLEEP12 powers down the DAC. PIN/SW and the actual hardware determine whether the applicable control is a pin or a register. The sleep pin in pin-control mode powers down the DAC. It is not equivalent to a software command that freezes phase accumulation.
With unchanged loading, capture the 50 MHz active oscillator at an appropriate MCLK point, the analog output at a DC-capable monitoring point, and a serial or sleep-pin marker. After an AC-coupled stage, a held level can decay toward bias and resemble a disabled DAC. With DC monitoring, preserve the necessary current-output termination and voltage compliance.
Choose a sine wave that makes the arithmetic clear. A 50 MHz active oscillator driving a nominal 6.25 MHz output gives a frequency ratio of one eighth, so each master cycle advances phase by forty-five degrees. Suppose the bench test establishes that accumulation was held for exactly 103 otherwise active cycles. At twenty nanoseconds per cycle, that is 2.06 microseconds. In that interval a continuous source advances 12.875 cycles, or 315 degrees modulo a cycle. The resumed output can therefore appear forty-five degrees ahead on the wrapped comparison. Use effective hold ticks, not the wall-clock interval between serial commands.
Test the two low-power controls separately
Keep frequency, loading, and control mode fixed. Pause the internal clock alone with the DAC enabled. Record held level, early recovery, and current. Restore the baseline and power down the DAC alone while leaving digital clock settings in place. Interpret the resulting analog level using the documented output and actual termination. Where the design exposes an independent digital observation, check that it can indicate continuing accumulation with the DAC off.
Distinguish freeze from reset. A freeze retains the previous accumulator state. Resetting the phase path requires evaluating a different start behavior. Begin the pause at two known phases and check whether held level follows the preceding waveform. If it always goes to one value, check for a reset or output control change. If the log shows a sleep write but the part remains in hardware control mode, interpret the actual active control.
Keep settled waveforms before and after each pause. The first cycles can reflect serial update timing, termination, and downstream filtering. For current comparisons, measure settled values with an appropriate slow instrument and record the transition separately with sufficient bandwidth. Avoid treating a transitional average as steady current. A DC baseline and the preceding analog sample help establish a held level instead of a changing waveform hidden by acquisition.
Compare recovery against a continuous reference
Capture a continuous coherent reference and the DUT output simultaneously. Record relative phase before stopping and after recovery. A separate freely running source can add ordinary phase drift to the comparison. Allow valid output to reappear if necessary, then document when the window starts. Do not use arbitrary visible edges as the baseline for repeated trials.
Also compare two pauses that differ by one actual master-clock tick, with the same prior phase relation. At the illustrated ratio, one extra held tick falls another forty-five degrees behind a continuous reference. Changing the software command duration may not change the effective number of frozen ticks. Log those counts, command timing, and recovered waveform separately. If frequency words, reset, or capture origin changed between runs, do not use this difference to infer the sleep effect.
After changing the reference oscillator, hold serial control and effective paused cycles fixed. If the held level and recovery phase follow the sleep control, verify the sequence before changing oscillator criteria. If MCLK itself stops, record external clock interruption and internal hold separately. Keep master-clock evidence, sleep mode, monitor location, recovery phase, and usable latency in the acceptance record.