Why does a 10 µs LTC6993 output stretch to 26 µs in a fixture timed by a 50 MHz oscillator?

A board fixture widens events for downstream logic. A single trigger gives about 10 µs high on the LTC6993, but a train of three keeps the output high for about 26 µs. The 50 MHz active oscillator still measures correctly. Determine whether each new trigger extends the existing pulse, and verify the full part suffix. Relative timing of the real input and output edges makes this difference easy to test.

Identify what sets the trigger timing and pulse width

ADI lists both the LTC6993-1 and LTC6993-2 as rising-edge devices. The first ignores events during its output pulse and subsequent rearming interval. The second can extend the active pulse if triggers meet its timing requirements. Falling-edge variants also come in both versions. The LTC6993 sets its basic pulse width using its own programming resistor and divider. The fixture uses a separate 50 MHz active oscillator to time the pulse generator. Event spacing comes from that source, whereas the monostable sets its own output width.

Start with widely separated single triggers. Suppose the measured basic pulse width is 10 µs with active-high output. Verify power, RSET, DIV, and the exact suffix. Probe TRIG and OUT under the expected load and check that both states of the stimulus meet the minimum input pulse-width requirements. A continuous 20 ns clock at TRIG would mix several timing limits. Use qualified, isolated pulses or a short burst with controlled spacing.

Add the next trigger before the output would fall

The example generator places three edges at zero, clock 400, and clock 800 from the 50 MHz oscillator. That gives 8 µs spacing. Each trigger can remain high for 1 µs for observation while leaving sufficient spacing. Neglecting small propagation differences, a retriggerable unit would initially fall at 10 µs. The second trigger moves that to 18 µs and the third to 26 µs. The output makes one rising transition and one falling transition. Use the tolerances on the physical unit when interpreting the illustrative total.

On a nonretriggerable device of the same polarity, the second trigger falls within the first output pulse and is ignored. The third arrives about 6 µs after the first pulse ends. If the rearming time and input conditions are met, it can start a distinct 10 µs pulse. Check whether the output stays up from zero to about 26 µs or falls near 10 µs and rises again near 16 µs. This directly verifies the signal downstream logic sees.

Follow with a two-pulse spacing sweep. Move the second edge from before the first output would end to after it. Check the minimum retrigger interval and the rearming requirement in the data sheet. Leave enough idle time between trials. Observe TRIG and OUT together, because a generator command does not establish the exact edge at the chip.

Confirm that the final valid edge sets the extended width

Move just the third trigger 1 µs later while it is still within the extended output pulse. On a retriggerable device that recognizes it, the falling edge should move about 1 µs too. Then remove it. The output should fall about one base width after the second trigger. This reversible test separates retriggered extension from the programming resistance, probe loading, and supply movement.

If an unexpected extension occurs only occasionally, look for reflection or pickup near the input threshold. Record TRIG and OUT in a single real-time acquisition with enough bandwidth for short glitches. Verify whether the extra edge was commanded or appeared on the interconnect. A digital decode can hide a brief event. If the change in falling time follows the programmed edge and the stimulus remains valid, the mode can explain it. An uncommanded pulse gives cause to inspect grounding, buffers, and edges.

A scope reading based on rising-edge frequency can be misleading while the output stays continuously high. Save the trigger IDs, output rise times, and final fall times together. If the downstream circuit treats a level as busy, check when it can accept work again, including during sustained bursts.

For another control run, space triggers farther apart than one output pulse plus the rearm time. The two variants can then look alike. Keep that sparse test as a baseline and include a short burst near the design spacing to tell them apart.

Record the minimum trigger separation, maximum continuous output width, and time to idle after the final recognized trigger over the required environmental and power conditions. A design that stays active through a train has a different acceptance test from one that requires a distinct pulse per event. When comparing 50 MHz active oscillators, hold the LTC6993 version, resistor, divider, and stimulus constant. If the pulse still stretches under the same rules, check the device suffix and event conditioning. Investigate the reference only if the actual spacing crosses a required boundary as the source changes.