Why do low-frequency points need more settling time in an AD5933 impedance sweep?

In an impedance sweep, the software may expect to wait the same amount of time at each frequency before reading the real and imaginary components. The AD5933 settling setting counts cycles of the current output frequency. A fixed cycle count therefore represents a different time in seconds at each point in the sweep. Insufficient settling can cause unstable early readings, especially when the device under test has appreciable reactance.

Convert the programmed cycle count to seconds

Suppose the setting uses 20 settling cycles. A 1 kHz stimulus then needs about 20 ms of pre-excitation, while 10 kHz needs 2 ms. This is only the settling before data collection. The full measurement includes acquisition, internal processing, and readout. Sum the times point by point using the actual frequencies to budget a sweep. If the master clock changed without updating the frequency settings, check the schedule against measured stimulus frequencies.

Run repeat sweeps with a calibration resistor first to verify that the small-signal chain is not clipping and that calibration coefficients apply to the band. Then connect the actual load, retaining excitation amplitude and gain. Increase the settling-cycle count at anomalous points and see whether both real and imaginary readings stabilize. If the resistor is repeatable but a capacitive load responds to the wait time, adjust the pre-excitation at those points.

Choose settling settings for repeatability and overall speed. High and low frequencies can use different durations and cycle counts. Record whether the instrument steps to isolated frequencies or sweeps continuously. If impedance phase is itself an acceptance criterion, record the measurement chain’s phase at the corresponding frequencies. One calibration point often cannot describe phase shift across a band. Also retain settling and calibration records for each gain setting on a multirange instrument.ADI AN-1252 on excitation settling and calibration