A test setup uses an AD650 to convert pulse rate to an analog voltage for long-term logging. After the source changes frequency, a counter may show the new rate while the voltage still transitions. Each valid input event contributes to the output, and the averaging network determines its response. Range, one-shot timing, and filtering affect the result.
Check the borrowed evaluation circuit against the manual’s frequency-to-voltage connection. Circuit values from the voltage-to-frequency example may not apply. Apply two stable pulse frequencies for long enough to settle. Confirm that one real edge produces one effective event, then measure steady voltage at each frequency to establish local slope and offset. Add a third point if needed to check linearity. The calibration applies to the tested range and settings.
Switch between the two frequencies, using the actual change in input edge timing as time zero, and record the output voltage. If the project specifies a maximum time to settle within an allowed voltage band, measure when the output enters the band and stays there. A crossing of the midpoint between the two steady levels measures something else. Heavier smoothing can reduce short-term voltage fluctuation when monitoring slow frequency drift. An overly long output time constant can hide a brief clock interruption.
Test clock interruptions of different durations to establish the detection limit. Note the recorder’s sampling and digital filtering. In acceptance records, separate steady-state conversion error, response time after a frequency step, and missing-pulse sensitivity. Incomplete analog settling can look like oscillator drift. Recheck the shortest detectable interruption if the filter time constant or recorder setting changes.ADI AD650 frequency-to-voltage circuit