An AD7705 weighing board still produces readings after a crystal change, yet its display moves around more than the old board when installed near a motor. Start with three measurements. What master clock is actually present? How often does DRDY announce a new result? How much does the output respond to a small injected signal near 50 Hz? The answers help distinguish a shifted filter notch from new interference related to grounding, common-mode levels or an overloaded input.
Identify the fitted source and read the four relevant settings
Read the clock register and record CLKDIV, CLK, FS1 and FS0. CLKDIV determines whether the master clock is divided by two before use. Table 19 uses 2.4576 MHz without division, or 4.9152 MHz with division, for the same internal clock. It specifies CLK at one in those cases. For 1 MHz without division or 2 MHz with division, it specifies CLK at zero. CLK affects internal current scaling as well as the rate selection. Updating a displayed sample-rate label alone does not configure the converter. For the divided external-frequency case, check the related conditions in the timing table.
Find out whether the board has a crystal connected between the two clock pins or a driven clock at MCLK IN. With an external driven source, CLKDIS can turn off MCLK OUT while the internal converter continues operating. With a crystal across MCLK IN and MCLK OUT, setting CLKDIS stops the clocks and conversions. When changing the circuit from one source type to the other, compare the pin connections and electrical requirements before checking DRDY again. Probe the oscillator using a method appropriate for that node, since placing a capacitive probe on a crystal terminal can alter the behavior you are trying to measure.
Use Table 20 to separate rate, rejection frequency and bandwidth
A notch is a frequency that the filter suppresses particularly strongly. In the AD7705 sinc-cubed digital filter, the first notch and output data rate have the same value, but they serve different purposes. The data rate schedules the reads, the notch helps reject interference, and the −3 dB point describes the low-pass bandwidth for useful changes in the signal. These examples use two of the manufacturer’s internal clock settings.
| Internal master clock | CLK / FS1 / FS0 | Data rate and first notch | −3 dB frequency |
|---|---|---|---|
| 2.4576 MHz | 1 / 0 / 0 | 50 Hz | 13.1 Hz |
| 2.4576 MHz | 1 / 0 / 1 | 60 Hz | About 15.7 Hz |
| 1 MHz | 0 / 0 / 1 | 25 Hz | 6.55 Hz |
Suppose the original board uses the first row for a 50 Hz first notch. A change to the 1 MHz internal clock and the last row gives a 25 Hz first notch. An ideal sinc-cubed response also has zeros at multiples of 25 Hz, so 50 Hz is the second one. The interval between updates grows from 20 ms to 40 ms, and the −3 dB bandwidth falls from 13.1 Hz to 6.55 Hz. A small shift of the interference or clock can move the alignment. Saturation ahead of the digital filter can change the result too. For a 60 Hz installation, the same 25 Hz configuration does not place a zero at 60 Hz. Consider the second row if its clock and bandwidth suit the board, or define separate filtering requirements.
Estimate the timing cost using an input step. The manufacturer gives a worst-case settling time of four output periods for a full-scale step. That is 4 ÷ 50 = 80 ms at 50 Hz and 4 ÷ 25 = 160 ms at 25 Hz. A system that must settle within 100 ms has already exceeded the allocation at 25 Hz before counting mechanical behavior or display processing. The data sheet also describes a special three-period case when the input step is synchronized using FSYNC to reset the filter. Apply that figure only to the relevant synchronized test condition. After changing the filter settings, initiate calibration as specified and distinguish its completion indication from an ordinary data update.
If the scale must also capture a 10 Hz change in load, add that point to the bench test. Using an ideal sinc-cubed envelope normalized at DC, the estimated loss at 10 Hz is 1.7 dB at 50 samples per second and 7.3 dB at 25. The analog input network, mechanics and any later averaging have their own responses. Inject the same 10 Hz amplitude in both tests and fit it using the recorded DRDY times to check the project's useful-signal budget.
Test rejection with a controlled low-voltage input
Keep the front-end gain, wiring, reference and supply fixed. Establish a suitable DC baseline and inject a limited low-voltage test signal near 49, 50 and 51 Hz through a coupling network appropriate for the circuit. Stay within the allowed differential and common-mode ranges. Measure the injected frequency and amplitude, then capture the raw conversion codes at DRDY. Save the noise floor with the source off and compare the response at the nearby frequencies and the step response after a setting change.
At 50 samples per second, residuals from 49 or 51 Hz can both fold to about 1 Hz. Residual energy when an injected tone coincides with the data rate can appear as an offset or a very slow beat. Looking for a 50 Hz FFT line, or subtracting the mean and then concluding the interference is gone, can miss this case. Compare with the injection-off baseline and use controlled relative phase or small, known frequency offsets to check the neighborhood of the notch.
Keep the full part numbers for the original and candidate clocks with the divider settings, measured DRDY interval, line-frequency suppression criterion, usable bandwidth and allowed settling time. Compare the boards in the same fixture. When the notch follows a change in update interval, investigate clocking and filter configuration. When the rate agrees but interference remains, check grounding, input common mode and overload before approving another component change.