Specialist Supplier of Crystals & Frequency Control Componentssales@mecxtal.com.cn0755-26500570
MERLIN home / Rednote / Measurement and Acceptance

Diagnosing a clock spur that moves when the analyzer changes its local oscillator

A spectral line appears in one receiver setting and shifts or disappears in another. Preserve both setups and verify actual carrier, reference, and resolution. Check receiver images as well. A superheterodyne receiver mixes RF with a local oscillator. Inputs equally spaced on either side can reach the same intermediate frequency. With limited preselection or rejection, a strong input can appear at an incorrect RF frequency.

Two RF signals can land at one IF

Take an illustrative single-IF receiver with a 10 MHz IF. With LO at 90 MHz, inputs at 100 and 80 MHz each sit 10 MHz away. If that setting interprets the response as upper-side input, the 80 MHz signal can appear at 100 MHz. They differ by twice the IF, or 20 MHz. To measure 100 MHz using a high-side LO at 110 MHz, the corresponding two possibilities are now 100 and 120 MHz. The true 100 MHz signal reaches the IF under either arrangement, but the 80 MHz signal no longer forms the same image there.

Actual analyzers may have multiple IFs, digital reception, and selectable preselection. Do not infer IF from display span. Consult the documented LO-side and image-rejection modes. Some modes compare several LO settings and take longer. Keep actual clock state stable when comparing, especially for sweeping or spread-spectrum devices.

Remove the suspected strong signal from the measurement path

Start with a broad span. If a strong auxiliary clock sits at 80 MHz, insert a characterized filter in the measurement branch and check loss at 100 MHz and rejection at 80 MHz. Suppose it adds only 0.5 dB loss at 100 MHz and 30 dB at 80 MHz. A corrected 100 MHz line that remains supports a real in-band component. A line dropping about 30 dB warrants checking an image. Filter changes can also alter overload or intermodulation, so preserve other levels and check the receiver's linear range.

Repeat using different LO orientation or image suppression. On an instrument without those controls, use preselection or a verified alternate receiver. Different attenuation and preamp settings can shift the floor. Hold verified input settings, check the carrier for compression, and record detectability. Save traces before averaging if two LO settings disagree.

Document the evidence at the actual limit

If a requirement limits a 100 MHz spur to −80 dBm, a −76 dBm result needs investigation when a second verified setup detects down to −90 dBm and finds no line. A second setup with only a −75 dBm floor provides little evidence either way. For burst outputs use the same observation and trigger timing to avoid mistaking off time for rejection. Retain the verification for each power measurement.

Report the strong input, suspected image, LO and preselection settings, attenuation, filter corrections, and floor. If disabling the suspected source alters board power or clock mode, first use a filter while leaving device operation unchanged, then compare source state separately. That distinguishes receiver behavior from board behavior.

If the bench permits, substitute a verified 80 MHz tone and vary its level while checking for an apparent 100 MHz response in the suspect LO setting. Restore the real device and verify it remains in the characterized power range. Some receiver changes occur only at path boundaries. Overlapping spans around a boundary can check reproducibility. A shared timebase does not rule out a receiver image. Export numerical trace data for weak lines so both settings can be compared with the same criterion.

References

WeChat

MERLIN · TIMING COMPONENTS

Scan with WeChat to contact Sales

Add our WeCom contacts from your usual WeChat app.

On desktop, scan with your phone. On mobile, save the QR image and select it from your album in WeChat Scan.