A small peak at 200 MHz does not directly establish the second harmonic of a 100 MHz clock. A strong fundamental can also drive the analyzer front end nonlinear and generate a response at the same frequency. The overlapping contributions show as one line. Its location alone cannot distinguish the source from the measurement setup.
Compare at different input attenuation settings
Hold the source, hookup, span, resolution bandwidth, and detector constant. Record the fundamental and second harmonic relative to it. After confirming the analyzer's input rating and DC limits, increase internal RF input attenuation, for example by 10 dB from a suitable initial setting. If the analyzer compensates display readings for attenuation, a stable external tone should remain nearly constant at the input reference plane. The signal reaching the internal mixer drops, changing distortion generated inside the analyzer.
If the relative peak changes substantially as attenuation increases, the original setup is affected by measurement-chain nonlinearity. Subtracting two screenshot values in decibels does not remove the internal harmonic. Device and instrument contributions at the same frequency can combine so the reading moves up or down. Try another suitable step to find a region where the reading is largely insensitive to attenuation and remains above the noise floor.
If the small peak disappears into the noise floor after increasing attenuation, report the upper level resolvable with that setup. To measure lower, place a characterized fundamental-rejection filter ahead of the analyzer and correct the path losses for the fundamental and second harmonic separately. State the output waveform. Harmonics of a digital square wave and distortion limits for a sine reference serve different purposes.Keysight procedure for identifying internally generated analyzer distortion