A network analyzer may show a jagged group-delay trace for a clock filter or interconnect. Group delay is the negative slope of transmission phase versus angular frequency. The instrument estimates the slope between phase points separated by a frequency span called the aperture. Ripple can come from the device or be amplified by phase noise when the aperture is too narrow.
Check the measurement with a known phase change
If unwrapped transmission phase decreases smoothly by 36° over 10 MHz, average group delay over that interval is about 36 ÷ 360 ÷ 10 MHz, or 10 ns. The same phase error produces more slope error over a narrower interval. Increasing the aperture can reduce measurement ripple but also averages genuine narrowband variation. For a narrowband filter, choose an aperture small enough to resolve the passband edge while retaining sufficient signal-to-noise ratio.
Keep the calibration plane, power, and IF bandwidth fixed. Repeat the same frequency range with two or three apertures and record the actual span in each case. Check sweep spacing and phase unwrapping so adjacent phase points do not jump excessively. In a deep stopband, weak transmission can make phase unreliable. A group-delay peak alone is insufficient evidence that energy remains in the network for that duration.
For a digital clock path, group delay at the carrier alone cannot predict when a square wave crosses the receiver threshold. Harmonics, amplitude response, and the threshold also matter. Report passband group delay separately from measured edge arrival. If you change the sweep point count, verify the aperture in Hz before retesting, because settings expressed as a percentage or number of points can change the actual span.Keysight group delay and aperture documentation