An occasional clock pulse reaches only half height. Should you use runt or pulse-width triggering?

A system occasionally misses a count when it changes clock sources, even though the average clock frequency looks right. A pulse that rises only halfway and a pulse that reaches normal height but drops too soon can leave the same symptom in the log. Start with an edge-triggered reference capture at the receiver. Note the normal voltage levels, pulse width and measurement threshold. Then run the same transition with separate runt and pulse-width triggers to determine whether the problem is amplitude, time at the high level or a missing edge.

Use two voltage thresholds to find a pulse that falls short

A positive runt rises through the lower threshold and then falls back through it without reaching the upper threshold. A negative runt falls through the upper threshold and returns without reaching the lower threshold. Select the expected polarity and put the two levels between the normal plateaus, guided by the receiver's limits and measured noise. An upper setting above the lowest normal peak can admit good pulses. A lower setting close to the low plateau or ground bounce can pick up noise. If the scope offers adjustable hysteresis, verify that the setting still catches the target pulse. If runt width qualification is available, start with any duration and narrow the condition after seeing the event. Keysight HD3 manual on positive and negative runt triggers, Rohde & Schwarz RTB 2 manual on runt thresholds and hysteresis.

Suppose a 1.8 V logic signal has a normal low no higher than 0.10 V and a normal high peak of at least 1.65 V. Set the lower trigger level to 0.55 V and the upper level to 1.20 V. A pulse that rises to 0.95 V and returns meets the positive runt condition. If the receiver requires at least 1.20 V for a guaranteed high at the operating supply and temperature, that pulse misses the guaranteed level. The actual switching point can lie within the undefined region, so the waveform alone does not establish what the receiver counted. A smaller disturbance that never crosses 0.55 V will escape this setup. To check for an entirely missing pulse, use timeout, same-slope edge interval triggering, or a control-line trigger and inspect the abnormally long period.

Measure short pulses at the receiver's timing reference

Suppose the receiver requires a high width of at least 4.0 ns measured at a 0.90 V reference. Set a positive pulse-width trigger and time the interval between the rising and falling crossings of that level. A rising crossing at 10.0 ns and a falling crossing at 12.8 ns give a width of 2.8 ns. If the pulse peaks at 1.70 V, it passed the 1.20 V upper threshold in the runt test but missed the width requirement by 1.2 ns. A runt that peaks at 0.95 V can also cross the 0.90 V width threshold. Check the saved waveform for height and ringing, then measure again at the reference specified by the receiver.

The 4.0 ns value is a receiver requirement. The trigger qualifier selects candidates and can include margin. If the trigger width error has been bounded at ±0.2 ns and normal pulses are much wider, trying a less-than-4.2 ns trigger gives room to collect borderline events for remeasurement. If a saved waveform measures 3.9 ns and that measurement also carries a ±0.2 ns uncertainty range, the interval from 3.7 to 4.1 ns overlaps the limit. Improve the timing measurement or use the project's specified decision rule. Check the scope's menu and trigger hardware separately. For example, the Keysight HD3 manual lists a minimum configurable width qualifier of either 2 ns or 5 ns depending on bandwidth model. Verify the shortest detectable pulse at the intended amplitude as well. For instrument acceptance, use a known test pulse close to the width and amplitude that matter. Keysight HD3 pulse-width qualification settings, Rohde & Schwarz RTO2000 trigger specifications by mode.

Capture the control action and the suspect pulse together

Wait for the event in Normal trigger mode or a single acquisition. Confirm the scope is armed before changing the clock source or enable. A runt condition may only be confirmed when the waveform returns through the first threshold, so leave enough time before the trigger to see the pulse start. Record enable, source select or reset on additional channels. If the relevant action leads the anomaly by about 6 µs, try 8 µs of pretrigger data and 2 µs after it. Capturing 10 µs at 5 GS/s needs about 50,000 points per channel. The sample interval is 0.2 ns, so a 2.8 ns pulse spans 14 intervals. Verify the effective rate once all required channels are active, and check whether the probe and channel bandwidth can resolve the edges. Sample count alone gives no width accuracy guarantee.

Save the first unaveraged waveform and the acquisition settings. Where probes allow, capture the driver output and receiver input in the same event. If the pulse is malformed at both points, follow the clock source, enable and switching circuit. If it leaves the driver intact and changes at the receiver, inspect loading, termination, connectors and ground close to the receiver. Check whether the probe capacitance or ground connection changes the edge. When comparing the order of transitions on different channels, account for relative probe and channel delay. Fast acquisition can help find rare waveforms, and segmented memory can retain many discrete events. Check the actual acquisition dead time, trigger rearm behavior and segment capacity. Tektronix 4/5/6 Series guide to fast acquisition and effective sample rate.

Count the switching events that were actually covered

Suppose two samples each go through 1,000 enable or source-switch events. For a fair comparison, verify that the instrument covered each relevant interval or record how many windows it did cover. A scope waiting only for an anomaly does not log every normal event. In the report, include the board and full part number, supply and temperature, receiver limits, trigger type and polarity, voltage and width thresholds, sample rate and event coverage. If the test yields no trigger, state that no event meeting the configured criterion was captured and identify any pulse types that could escape it. If an event is captured, provide the unaveraged receiver waveform with the control signals from that same acquisition so quality staff or the supplier can repeat the test.