Specialist Supplier of Crystals & Frequency Control Componentssales@mecxtal.com.cn0755-26500570
MERLIN home / Rednote / Measurement and Acceptance

Two lines crowd a 10 MHz carrier. How should you choose analyzer resolution bandwidth?

A 10 MHz oscillator shows a shoulder roughly 2 kHz to the right of the carrier. Zoom in and the shoulder is still joined to the carrier. Save the wide-span trace, then check the analyzer's actual resolution bandwidth (RBW), measurement mode and input level. To decide whether there is a distinct spur, improve the frequency resolution and see whether a separate peak repeats at the same offset and level.

Use the expected 2 kHz spacing to set span and RBW

Suppose the carrier is at 10.000 MHz and the candidate at 10.002 MHz. Their separation is 0.002 MHz, or 2 kHz. Center the measurement near the observed carrier and try a total span of 20 kHz. At a 10.000 MHz center, that covers 9.990 to 10.010 MHz, including the +2 kHz and −2 kHz offsets and background on both sides. Save a coarse trace, then repeat at available RBWs near 500 Hz and 100 Hz. Two continuous tones of similar level may separate as the filter narrows. If one is much weaker, the carrier's filter skirt or phase noise can still mask it even with a 500 Hz RBW.

On a swept analyzer, RBW commonly means the −3 dB bandwidth of the resolution filter. The ratio of −60 dB bandwidth to −3 dB bandwidth helps describe its selectivity. What matters for the weak tone is how much stronger signal leaks through the filter 2 kHz off center, and how much phase noise is present there. A nominal RBW smaller than 2 kHz gives a starting point. If the candidate still blends into the skirt around the carrier, report it as unresolved. Keysight Spectrum Analysis Basics on selectivity and close-in phase noise.

Change RBW first, then test for distortion inside the analyzer

In the first comparison, keep the input attenuation, preamp, reference level, video bandwidth (VBW), detector and averaging fixed where the instrument allows. Let the analyzer choose a suitable sweep time for each RBW, allow it to settle, and record the actual time. In a traditional analog sweep at fixed span, the required time scales approximately with 1 / RBW². In that model, reducing RBW from 1 kHz to 100 Hz can take 100 times as long. Digital filtering and FFT-based processing follow their own timing, so use the actual instrument setting. Sweeping too quickly can reduce a stable tone's displayed amplitude or move its indicated frequency. If VBW is automatically linked to RBW, fix it where feasible or record both actual values for every trace. Keysight on sweep time and RBW, Rohde & Schwarz FPC manual on RBW and VBW.

For locally flat noise and filters of the same shape, reducing effective noise bandwidth by a factor of ten lowers the unnormalized noise power by about 10 dB. A properly measured continuous tone should hold about the same peak level once the trace settles. Compare like units. Marker readings in dBm differ from values normalized to 1 Hz in dBm/Hz or dBc/Hz. As an example, a locally flat noise level of −100 dBc/Hz near +2 kHz contributes about −80 dBc in 100 Hz of equivalent noise bandwidth. With a strong carrier, the source and the analyzer's local oscillator can both contribute to the close-in background. Rohde & Schwarz on phase noise and bandwidth normalization.

For a second comparison, keep the selected RBW and increase the input attenuation by a known step, such as 10 dB. Ensure the input has sufficient margin against overload and the candidate remains above the floor. If the input-referred level stays about the same, that supports a component already present at the input. If it changes substantially, check for internal distortion or overload. If added attenuation merely buries the candidate in the noise, the test is inconclusive. You can also substitute a known low-spur, low-phase-noise 10 MHz reference at a comparable input level and repeat the sweep. That checks what this setup can resolve close to a carrier. Keysight's input attenuator test for analyzer-generated distortion.

For an FFT, lengthen the real record and check the window

On an oscilloscope FFT or vector signal analyzer, a record of duration T has a native bin spacing of 1 / T before zero padding. A 0.5 ms record gives 2 kHz bin spacing. Keysight 89600 VSA defines its ResBW using the window's equivalent noise bandwidth. Under that definition, a Hann window gives about 1.5 / T, or 3 kHz for this record. Extend the actual record to 15 ms and the bin spacing becomes about 66.7 Hz with about 100 Hz equivalent noise bandwidth. Zero padding can produce a finer plotted grid without adding captured time. Some instruments report a −3 dB bandwidth and others report equivalent noise bandwidth. Record the definition when comparing instruments. Keysight 89600 VSA ResBW definition, Keysight on window bandwidth, shape and sidelobes.

A strong carrier can also leak into neighboring FFT bins. Keep the window fixed while collecting a longer record and see whether the peak remains at the same offset. If you try a lower-sidelobe window to reduce leakage, record the new window, time length and RBW. The main-lobe width and amplitude correction can change as well. Repeat the measurement if a candidate appears or disappears with changes to padding or bin selection alone. If the line is genuinely intermittent, save records around the event and use a time-resolved spectrum to correlate it with the operating state.

Give the supplier a repeatable offset and level

Confirm the oscillator output type and intended load before connecting it. Use a suitable buffer, termination and attenuation. The analyzer's 50 Ω input can change the behavior of an output designed for a high-impedance CMOS load. If you use a buffer or DC block, check that 10 MHz is in its passband and that the resulting input level and DC offset are within the analyzer limits. Keysight 3588A example of connecting an oscillator and selecting input impedance.

On each capture, identify the interface and test point, carrier level, signed offset and spur level. For example, a −12 dBm carrier and a −80 dBm candidate give −68 dBc. Lines at matching positive and negative offsets can be reported as possible related sidebands. To attribute them to supply modulation, frequency modulation or another periodic disturbance, compare with the supply or operating state, or use a modulation measurement.

Include center, span, actual RBW and VBW, detector, averaging, input attenuation and sweep time. For an FFT, add the window and actual record length. Save the neighboring background and identify the supply, load, temperature, full part number and lot. If the candidate remains indistinguishable from the background or carrier skirt, report it as unresolved at +2 kHz and state the measurement's available dynamic range. Compare against a supplier limit once the setup can resolve a signal of that level at the stated offset.

WeChat

MERLIN · TIMING COMPONENTS

Scan with WeChat to contact Sales

Add our WeCom contacts from your usual WeChat app.

On desktop, scan with your phone. On mobile, save the QR image and select it from your album in WeChat Scan.