Detecting a 1 Hz weekly drift when every daily frequency-station check passes

Each day an incoming inspection station logs a check device. All observations remain within the individual control limits, but a week of results moves in one direction. The source could be the bench, reference, fixture, or check device. An exponentially weighted moving average can flag persistent movement using a stable historical baseline. Its control limits describe the measurement process. Product disposition still follows the product specification and decision rule.

Choose a check device that exercises the real method

Use an independently measured, repeatable oscillator and document its serial number, storage and warm-up conditions, load, socket, supply, and temperature. Insert it into the measurement chain that tests product. An external source feeding the counter reference checks part of the timebase path. A check at the product position can also reveal load changes and worn contacts. An active source can bypass excitation and load circuitry in a passive-crystal fixture.

Build the initial baseline from a stable period and cover ordinary shifts, sockets, and environment. Keep records of investigated events. When replacing an aging check device, compare the old and new units during a transition before setting a new baseline. If a check device experienced thermal shock or rework, use an independent reference and a backup device to determine which quantity changed.

Small shifts can accumulate in the chart

Assume a historical mean of 0.0 Hz and single-reading process standard deviation of 0.4 Hz, with a common frequency zero. Under a simplified independent stable error model, three-standard-deviation individual limits are ±1.2 Hz. Five daily readings of +0.2, +0.4, +0.7, +0.9, and +1.0 Hz all remain inside. Start the EWMA at the baseline and update with 0.25 times each new observation plus 0.75 times the previous EWMA. The values are about 0.050, 0.138, 0.278, 0.434, and 0.575 Hz.

The fixed steady-state three-standard-deviation EWMA limits are ±3 × 0.4 × √(0.25/1.75), or about ±0.454 Hz. With those fixed limits, day five calls for investigation. Startup-adjusted limits can alert on a different day. Use stable historical data and acceptable false-alarm rates to set real weights and limits. Assess autocorrelation, thermal cycles, or shift-specific differences before choosing the chart. Treating the suspect week as a baseline would mix the drift into the estimate.

Investigate the time window and affected units

Preserve the sequence leading to the signal. Hold potentially affected results under the procedure and identify the last verified bench state, fixture changes, and lot timing. Repeat the check under matched conditions with the original record intact. Cross-check the reference, counter, and fixture independently. If the check device moves by 1 Hz on another bench, investigate aging and storage. If a backup device reads high only at the original station, prioritize that bench. Record any changed supply or temperature mode instead of silently resetting the plot.

If maintenance introduced a +1 Hz bias, prioritize review of products near their decision limits. The review scope also depends on uncertainty in when the bias began, measurement rules, and possible impact. After repair, compare a verified check device with an independent reference and demonstrate control before releasing the station. Reset the baseline only with a documented rationale such as an approved method change.

Keep the raw error in hertz, reference zero, weight, actual check timestamps, and fixed limits in a reproducible record. Mark missing days explicitly. When stations share a reference, compare when shifts began rather than averaging unrelated timelines. Use that evidence to bound the affected bench, check device, and production records.

If a station takes three readings per day, define whether the chart uses the first, the mean, or individual readings. If it uses means, estimate historical deviation from means formed the same way. The 0.4 Hz single-reading deviation does not automatically apply. Retain all three values to investigate changes in within-session repeatability. If a production run delays a check until night, verify that supply and environment are stable and apply the documented rule for the delayed reading.

References