Specialist Supplier of Crystals & Frequency Control Componentssales@mecxtal.com.cn0755-26500570
MERLIN home / Rednote / Measurement and Acceptance

Did divide-by-eight triggering open a 500 Mb/s eye by leaving out part of the pattern?

When an external trigger input cannot take the full bit clock, a divided clock can produce a stable eye. A cleaner display may come from incomplete pattern coverage. If the repeating pattern length and division factor share a divisor, triggers can revisit only a subset of its bit positions. Verify coverage before comparing sources, terminations, or drive settings.

Where does every eighth trigger land?

Let L be the repeated pattern length and D the clock division ratio. Each trigger advances D positions modulo L. The sequence repeats after L divided by the greatest common divisor of L and D. If the scope folds only the one UI after each trigger, that quotient is the number of distinct observed positions. With L = 32 and D = 8, triggers land at 0, 8, 16, and 24, then return to 0. The other 28 positions are absent from that view. More triggers cannot reveal a late transition that always occurs at bit 9.

With a 127-bit PRBS7 and D = 8, the values are coprime and ideal triggering visits all 127 positions. Actual acquisition still needs sufficient captures and must not skip a patterned subset during rearm. A scope folding many UIs per record covers additional positions. Read the actual record length and folding algorithm before applying the one-UI example.

Verify positions against a pattern marker

Enable a marker identifying the start of the test pattern, or align raw data to the known sequence. Capture at least a complete pattern with sufficient sampling per UI. At 500 Mb/s one UI is 2 ns and 32 bits take 64 ns. A 10 ns segment cannot establish whole-pattern coverage by itself. Compare crossing times by bit position over a longer record or stepped measurement window.

For one-UI acquisitions, use an available trigger phase or pattern position control to reach distinct residue classes without altering the transmitted edges. Compare eye crossings in included and previously missing positions. Suitable full-rate triggering, pattern sync with a long record, or the instrument's pattern-aware acquisition can also provide coverage. If using recovered-clock triggering, record the loop settings. Tracking slow phases and missing pattern positions are separate effects.

Judge the eye with its actual coverage

Suppose one view reports 1.5 ns and a complete-pattern capture reports 1.2 ns. At 2 ns/UI these are 0.75 and 0.60 UI. For a 0.65 UI minimum, the complete capture warrants review under equivalent levels, loading, and algorithms. If the load also changed, isolate that variable. Preserve crossing measurements by position and the raw examples of the worst edges.

Report pattern length, division ratio, trigger source and phase, UIs per record, and evidence that all intended positions appear. Record persistence and valid acquisition counts. A long BER run and a short scope capture measure different observation intervals. Establish eye-pattern coverage before correlating it with errors or crediting a hardware improvement.

Validate the menu with a deliberately constructed 32-bit sequence. Place the first transition after a long run of zeros in a suspected missing position. Compare full-rate and divided-trigger captures after verifying full-rate trigger bandwidth and timing. Some generators insert idle, synchronization, or control words after a PRBS, so the true repeating sequence can be longer than the stated PRBS length. Include transitions relevant to normal operation.

Tabulate valid acquisitions by pattern position. Zero counts focus attention on triggering or folding. A few observations at one position versus thousands elsewhere can also be insufficient for the specified statistics. Save the coverage counts alongside the eye so another bench can verify whether it observed the worst pattern.

References

WeChat

MERLIN · TIMING COMPONENTS

Scan with WeChat to contact Sales

Add our WeCom contacts from your usual WeChat app.

On desktop, scan with your phone. On mobile, save the QR image and select it from your album in WeChat Scan.