Why a 100 GS/s oscilloscope can still miss the first oscillator edge after power-up

Zooming in on a clock edge can make an acquisition look more convincing. On some instruments the highest advertised sampling rate belongs to equivalent-time operation. It applies to waveforms that repeat and can be aligned. When a test concerns one power-up, its first edge, its first complete cycle, or an intermittent missing pulse, establish whether the trace came from a single acquisition. Combining points from different starts can produce a shape that never occurred on any individual start.

What the two sampling rates mean

A real-time acquisition collects its entire record in one trigger-associated acquisition. Equivalent-time sampling collects points across many triggers and reconstructs a waveform using their positions relative to the trigger. Some instruments can select a mode automatically. Read the selected mode, actual sample rate, and active channel configuration from the scope and the waveform file. The headline rate usually assumes a particular mode and channel configuration.

An illustrative scope delivers 2 GS/s in real time, or one sample interval every 500 ps. Its 100 GS/s equivalent-time grid is spaced at 10 ps. A 1 ns edge spans about two real-time sample intervals in any one record and about a hundred intervals on the reconstructed grid. To obtain genuinely more information about one edge, increase the real-time rate, reduce the number of enabled channels where supported, or use a scope with sufficient real-time rate and bandwidth. Keep track of which points came from interpolation.

Repeated starts may disagree

Imagine triggering ten starts from the power-good signal. Five yield a first rising edge at 40 ns and the other five at 47 ns. Each rise takes 1 ns. If equivalent-time samples at 44 ns came from several starts, some would show a high level and others a low level. The display might show a broad or doubled edge depending on acquisition and rendering. That picture cannot establish the rise time of one start or whether each first cycle was complete.

If just one start contains a 2 ns first pulse, equivalent-time triggering on the settled clock favors its repetitive later edges. Triggering on power-good still leaves the reconstructed points drawn from different starts. Repeated power cycling remains useful if each cycle has a saved real-time record and the scope is ready when that cycle starts.

Capture the sequence first and the edge second

Start with a long record. Capture the start request or power-good on one channel and the clock on another, with enough pretrigger data to establish event order. Recheck real-time rate if adding a channel for output enable. Reproduce the initial conditions, including power-off interval and residual voltage. Document whether timing starts at the supply, power-good, or enable. Keep the native acquisition file so interpolated or averaged exports can be traced back to original samples.

Use a second, shorter acquisition near the expected first edge to raise the real-time sample rate. Measure the first valid high and low intervals and the point after which valid pulses are continuous. Millisecond startup observations and nanosecond pulse measurements can use two settings with matching power-off conditions, load, supply, and probe location. If the long record found an anomaly, a normal short record from a later start belongs to that later start. Preserve the abnormal capture and its time tag.

What belongs in the acceptance record

An illustrative rule allows 50 ns from power-good to the first complete cycle meeting the receiver's level limits and minimum high and low widths. The procedure also needs to define the edge that marks cycle completion. Actual limits come from the part and interface. Save both waveforms, the sample rate, the trigger source, and the timestamp for each run. Check raw point coverage around threshold crossings on marginal samples and corroborate with a higher real-time rate when necessary. Equivalent-time instruments can still characterize repetitive edges in steady state. Identify which acquisition produced each result.

Some instruments also indicate how many acquisitions contributed to a trace. Arm a single acquisition and allow only one power cycle. If the scope cannot render that start in full, check the acquisition mode and record before diagnosing the board. When starts repeat reliably, compare the two acquisition modes in the settled portion as a sanity check. If the steady-state shapes disagree substantially at the same point, use a known fast edge to verify channel and fixture settings before comparing starts.

References