A 32.768 kHz tuning-fork crystal meets its requirement at 25 °C but runs slightly faster after the temperature rises a few degrees. This alone does not establish a fault in the chamber or counter. For the Epson FC2012AA ordering code X1A0001810003, 25 °C is the initial frequency-tolerance measurement point, while the turnover temperature is specified as 25 °C ± 5 °C. The specification allows the room-temperature calibration point to differ from the peak of an individual crystal's frequency–temperature curve.
Use a measurement circuit with the 12.5 pF load specified for this order code, and keep excitation and supply constant. Count at a buffered output to isolate the resonator node from cable capacitance. Take readings at several temperatures on both sides of the expected maximum after local temperature and frequency stabilize. Using different measurement circuits at the warm and cold ends can introduce parasitic differences that look like crystal drift.
Convert each reading to deviation from the same frequency reference. Plot deviation versus temperature and fit a quadratic near the maximum. The slope is small near the peak, so the frequency difference between adjacent points can be smaller than measurement repeatability. Keep repeat measurements and fit residuals to determine how many reported digits the temperature increments support.
Record the 25 °C deviation, fitted turnover temperature, and original data on both sides of the maximum. For equipment timekeeping at a seasonal temperature, use the deviation at the actual operating point. The room-temperature result does not apply throughout the rated range. Confirm the complete order code before comparing another load option in the same family.Epson FC2012AA static characteristics for the specified order code