A board scans four capacitive electrodes. After changing the reference, the application expects quicker updates and sees unstable results around channel switches. The FDC2214 measures the oscillation frequency of an LC sensor, using a reference to time activation and conversion. The same count values represent less physical time at a faster reference. The result can arrive earlier while the chosen sensor network has less time to settle.
Convert each count separately
TI describes SETTLECOUNT for sensor activation and RCOUNT for conversion. Use the actual effective channel fREF after any reference division. At an assumed effective 20 MHz, a SETTLECOUNT of 500 times sixteen reference cycles is 400 µs. An RCOUNT of 2,500 gives about 2,000.2 µs using the detailed equation with four added clocks. At 40 MHz with unchanged values these become 200 µs and 1,000.1 µs. Channel-switch timing is additional.
To retain approximately the same durations with the faster clock, begin the calculation around SETTLECOUNT 1,000 and RCOUNT 5,000. Conversion then takes about 2,000.1 µs using the detailed equation. Verify device conditions for reference, dividers and sensor input frequencies. Choose actual counts according to LC startup and required resolution. If retaining measurement quality requires the same durations, throughput can remain near the previous value despite the faster crystal oscillator.
Use differences between channels to isolate settling
Electrode and cable construction change sensor frequency, loss and startup. Round-robin switching exercises settling conditions that continuous sampling of one channel may avoid. Record settled readings, first-sample deviation and time on each channel. Hold RCOUNT fixed and extend the slow channel's settling first. If that resolves its first readings, keep activation margin there. If settling makes little difference but increasing RCOUNT reduces spread, evaluate the conversion time and update-rate trade. Changing both together obscures the cause.
Four channels at 0.4 ms activation and roughly 2 ms conversion require about 9.6 ms per scan before switching overhead. Each channel gets one result per scan, so the 2 ms conversion does not mean 500 updates per second per input. Calculate cadence from the configured sequence if channels are sampled unequally. Label per-channel and aggregate rates. A host reading its management bus faster can retrieve the same completed result repeatedly. Count fresh results by the applicable status or notification.
Retain clock and detection evidence
Use stable reference inputs or fixed capacitances to compare clock changes, channel changes and cable changes separately. Also retest across actual installation cable length and grounding. Save the reference waveform and configuration used to derive fREF. A frequency reading at the oscillator is not sufficient if the channels divide it. Keep fREF, dividers, activation and conversion budgets, status flags, code spread, and first values after switching. Capture low-rate triggers and cold starts separately from the continuous result. Report the measured update interval per channel and the slowest sensor settings.
Preserve a capture from the previous board. Hold the electrode and target geometry fixed and compare only reference and corresponding register settings. Observe startup and settled codes separately. If geometry cannot be held constant, compare against a repeatable electrical reference first and use a separate installation test.
Swap two electrodes at the input pins while keeping the device’s preset scan order. If unstable readings follow one sensor, examine its activation, cable and installation. If the first result after switching is always affected, inspect switching delay, data-ready handling and channel mapping. Keep channel identifiers and actual read times with raw values to avoid mistaking the scan cadence for sensor variation.