A 25 MHz device allows ±10 ppm, or ±250 Hz. The reading is +249 Hz, within specification by only 1 Hz. If expanded uncertainty for the actual measurement is 5 Hz, the inspector needs to know whether the agreement uses the indicated value directly or applies a guard band. A counter with more display digits cannot decide that question.
Record specification limits separately from acceptance limits
Suppose the parties agree to tighten both specification limits by 5 Hz for direct acceptance. The upper acceptance boundary is then +245 Hz, and a +249 Hz reading enters the agreed review region. That does not establish that the true frequency is outside specification. Simple acceptance may give another result. Agree on guard band and borderline treatment before measuring.
Use uncertainty for the actual conditions, including reference timebase, temperature, loading, reading method and repeatability as applicable. Display resolution alone does not establish uncertainty. Repeating a reading ten times cannot remove a systematic error from the common reference.
Keep the unrounded result, calibration status and uncertainty for borderline parts, then perform the retest or higher-accuracy comparison specified by the rule. Put the device limits on the order and the acceptance rule and review path in the inspection procedure. Both parties can then make the same decision at +249 Hz. If external reports use rounded values, keep the agreed precision for internal decisions.Keysight guidance on guard bands and measurement decision rules