To isolate a fault, an engineer may remove the original driver and inject an 8 MHz signal into the receiver's external clock input. The generator displays 1 Vpp, yet the IC can see nearly 2 Vpp. Check the generator's expected-load setting. Some generators have a fixed 50 Ω source impedance and calculate displayed amplitude and offset for the selected termination.
Calculate the voltage division before connecting the pin
In the ideal low-frequency model, a 50 Ω source in series with a 50 Ω load delivers half the open-circuit voltage to the load. If the generator displays 1 Vpp for a 50 Ω termination but actually drives a high-impedance input, the voltage approaches the 2 Vpp open-circuit value. DC offset can also differ. Changing the expected-load menu from 50 Ω to high impedance may alter displayed amplitude and offset on some instruments. Check the operating guide for whether it changes the physical output.
Confirm which receiver pin permits an external drive, its normal operating levels, and its absolute maximum range. Disable the board's existing output as instructed by the device manufacturer to avoid two drivers operating at once. Before connecting the IC, verify the waveform with the same hookup and an equivalent load. If another instrument with a 50 Ω input is attached to the same node through a tee, that termination will also affect the voltage at the pin. For the substitution trial, compare the high and low levels and overshoot at the receiver pin itself.
With longer connections, reflections make instantaneous edge amplitude differ from simple voltage division. Verify at the receiver and account for probe loading. Record generator source impedance, load setting, fitted termination, offset, and measured pin voltage so the same injection conditions can be reproduced on another setup.Keysight function generator output load settings