A pulse-per-second reference can itself have quantization error. In the u-blox 6 example, a finite-rate internal clock generates pulses, so the ideal second does not always coincide with an available transition. The module reports the quantization error for a particular pulse. Pairing it with the wrong pulse can increase jitter.
Confirm model, pulse rate and message semantics
The application note describes quantization correction for 1 Hz and the relationship of UBX-TIM-TP to the next pulse. Pair messages using protocol time tags, validity and units rather than host serial arrival order. For newer modules or firmware, use their matching interface description.
During one continuous capture, retain raw pulse error against reference, message timestamp, correction and validity. Generate uncorrected and corrected plots from the same data. If scatter grows, check sign, units and whether pairing is one second off. Mark dropped samples as gaps instead of shifting all subsequent messages.
An independent fixed delay remains in the mean, so lower short-term scatter does not establish absolute alignment. In TCXO acceptance, state reference limitations and matched sample count, and retain raw pulse records for recalculation after decoder changes. For pulse rates above 1 Hz, check what corrections that mode provides.u-blox 6 GPS timing application note, Section 1.2