A 10 MHz reference brings its one-second pulse and RF output to test headers. An inspector sets both oscilloscope channels to 50 Ω. Whether that supports an amplitude decision depends on each output's conditions. The IQD IQCM-200 illustrates an easy-to-miss distinction: outputs from one module can have different specified loads.
Draw each output separately on the test sheet
Page 2 specifies 50 Ω for the one-second pulse output levels. The HCMOS 10 MHz output specifies 15 pF, and a separate 50 Ω sine output option is listed. “10 MHz plus 1PPS” is insufficient. Confirm the RF output from the actual order, then record waveform, load, and acceptance criterion for each test point.
For the HCMOS RF option, dividing a 2.7 V high level by 50 Ω yields 54 mA. That is the current an ideal resistor draws at that voltage. The manufacturer's 10 MHz, 15 pF condition does not guarantee that DC drive. A small amplitude under the wrong load does not establish sample failure. Remove the incorrect termination, restore the specified load for the ordered version, and capture again.
The input is another easy line to misread
The same page specifies logic thresholds and test conditions for the 1PPS reference input. This receives an external reference and does not use the output row's test method. Mark signal directions and identify the generator, device under test, and oscilloscope on the hookup drawing so two identically named pulse outputs are not wired together.
At receipt, retain the module label, confirmed output option, channel settings, and captures. If the cabinet originally used a 50 Ω sine reference, specify that option for spares. Matching the data sheet's cover model and nominal frequency can still result in a functioning module incompatible with the existing harness.
IQD IQCM-200 data sheet, page 2: outputs and load conditions