A passive probe clipped to an end of the frequency-setting resistor alters the previously stable output of an LTC1799 clock board. SET participates in frequency setting. The manufacturer limits added capacitance there and recommends locating the resistor close to the device. A probe on that node changes the circuit under test.
First record the frequency while observing only OUT and leaving other conditions unchanged. Look up the probe input capacitance and estimate the contribution of the test pad, trace, and fixture. The manual recommends less than 10 pF on SET. A conventional passive probe alone can use much of that allowance. This recommendation applies to SET. The output load capacitance is a separate specification. To confirm the setting resistor, inspect soldering and measure resistance with power removed where appropriate.
Compare with a lower-capacitance measurement method and record whether frequency returns to baseline when the probe is removed. A remaining error warrants reviewing resistor tolerance, DIV, supply, and layout. If the error appears only with the fixture attached, redesign the test contact and obtain production frequency readings at a low-impedance output where possible.
Document permissible probe points, probe model, and expected frequency variation for service. This prevents a test of the setting network from making each board run under conditions different from the customer’s. If an in-circuit test point must remain, requalify frequency with the normal fixture before retaining its pad and trace.ADI LTC1799 SET pin and board-layout requirements