Calibration equipment uses a 10 MHz OCXO. The old Microchip OX-171 configuration uses a 5 V supply and sine output. A board revision moves to 3.3 V and proposes a part from the same family. Both offers say 10 MHz and 50 Ω, but reusing the old attenuation network can leave a different receiver power budget. The manufacturer gives different sine-output power limits by supply grade. Ask engineering to include the actual attenuation and receiver range before approving the replacement.

Use the sine power limits for the ordered supply

In the OX-171 RF table, the 3.3 V option gives 2 dBm minimum, 5 dBm typical, and 8 dBm maximum into 50 Ω. Its 5.0 and 12 V options give 5, 8, and 11 dBm respectively. Supply codes E, D, and B designate 3.3, 5, and 12 V, while an E in the output field indicates sine. Record both fields and the complete number. The two E positions represent different options. The 12 V supply has restrictions on temperature combinations, so confirm the requested configuration. The sheet also lists HCMOS electrical conditions. Keep them separate from the 50 Ω sine-power requirement.

A dBm value is referenced to 1 mW. The minimum 2 dBm is about 1.585 mW. At 50 Ω, take the square root of power times resistance to get about 0.282 V RMS. For a sine wave that is about 0.796 V peak to peak. The 5 dBm minimum for the 5 V version corresponds to approximately 1.125 V peak to peak under the same load. This illustrates how a supply-grade change can reduce receiver margin. A high-impedance scope connection does not reproduce the voltage under the specified 50 Ω load. Confirm impedances before accepting either power or loaded amplitude.

Check both ends of the receiver power window

Suppose a board requires 4 to 9 dBm at the receiver and its 10 MHz path has 1 dB of loss, taken as constant for the example. The 3.3 V version gives roughly 1 to 7 dBm at the receiver, with a 3 dB shortfall at the low end. The 5 V version gives 4 to 10 dBm, 1 dB above the maximum at the high end. Both typical figures look usable. The receiver window spans 5 dB while the source range spans 6 dB. Shifting the entire interval with a fixed attenuator cannot make it fit. Review a wider input range, appropriate conditioning, or a controlled source amplitude with the actual board evidence.

Use the actual losses of the connectors, cable, board network, and any attenuator. A short cable at inspection may omit the enclosure routing and splitters. Measure at the specified oscillator load and at the receiver. Use compatible calibration and reference cables, noting adapters and any loss limits used in the budget. For low receiver power, check the supply grade and full code and confirm source performance first. Then measure the actual installed link. A measurement at an inapplicable receiver port cannot support rejection of the whole incoming lot.

If the installed path has a range of losses, use both extremes. Suppose a short measured path shows 0.8 dB while the project allows up to 1.4 dB for connectors and routing. For the 5 V source, subtract maximum loss from minimum source power to get 3.6 dBm. Subtract minimum loss from 11 dBm to get 10.2 dBm at the high end. Allow for temperature, mismatch, and measurement uncertainty as applicable. A short test cable cannot represent the longest production path without that comparison.

For a split output, identify whether the measurement point is before or after the splitter. When moving one instrument between outputs, disconnecting another receiver can change the loading. Record whether both receivers were operating. Reproduce the configuration with approved equivalent loads, then verify the installed equipment. Retain port states with the result so the next lot can use the matching board revision.

Retain the output conditions from quote through repair

Specify supply, sine output, minimum and maximum power, test load, and complete code for the 10 MHz OCXO. The work order can also state receiver limits, path loss, and buffering for board approval. Receiving checks identity, supply, orientation, and frequency and then measures power under the proper load. Verify the installed receiver level separately. If the supplier provides only a typical figure or family overview, obtain the applicable commitment before approving the part for the existing budget. A frequency-only station can support an identity check while loaded output remains unverified.

Use these records for a cable change, receiver revision, or new supply. Preserve the real order code and its board applicability when visually similar 10 MHz parts are stored together. If a receiver fails to lock despite a matching nominal frequency, use an approved sine-output part for comparison and check power at both points. That establishes whether the oscillator meets its purchase requirement and whether the assembled path matches the approved conditions.