A test instrument shuts down each day. After repair, the team wants to know how closely its reference returns to the old frequency when power comes back. Microchip's OX-176 data sheet lists output start time, warm-up time and frequency retrace after a power interruption. A lab request that simply says “measure frequency after power-on” could produce a reading at the first visible RF output, after five minutes, or the next day. For this 10 MHz OCXO, each observation answers a different acceptance question. Establish whether the system needs early RF amplitude, thermal settling, or frequency repeatability after a complete off interval.

Define retrace with two frequency readings and one power interruption

The OX-176 retrace limit is ±10 ppb. The note calls for a reference reading after 24 hours powered, followed by 24 hours off and a second reading after 60 minutes powered again. Subtract the reference reading from the second reading. If the old equipment had an OX-174 from the same manufacturer, consult that part's own sheet. Its note compares the end of two 24-hour powered periods with a 24-hour off period between them and constant temperature. A generic “power-cycle retrace” shortcut does not state whether to read at one hour or 24 hours after restart. Put the oscillator model, specification and three intervals on the purchase attachment.

Suppose the original powered reading was 10 MHz plus 0.18 Hz, and the 60-minute reading after restart was 10 MHz plus 0.27 Hz. Their difference is +0.09 Hz. At 10 MHz, one ppb is 0.01 Hz, making the retrace +9 ppb and within ±10 ppb. The restarted output can still be 27 ppb above nominal. If another reading after 24 hours powered again is 10 MHz plus 0.30 Hz, the change from the original baseline is +12 ppb. That later reading can describe subsequent drift. It does not replace the specified one-hour result.

Warm-up has its own reference

Microchip gives a maximum five-minute warm-up at 25°C to within ±10 ppb of the one-hour final frequency. The reference belongs to the same power cycle. If the one-hour value is 10 MHz plus 0.27 Hz and the five-minute value in that cycle is 10 MHz plus 0.36 Hz, the warm-up difference is also 9 ppb. This does not supply the missing cross-cycle retrace measurement. The RF table separately gives a typical two-second start time to reach 90% amplitude. An instrument that starts using the reference as soon as it sees a waveform needs its own early-use amplitude and frequency evidence.

A lab may warm several units at once and switch one instrument among them. If the switch removes power or changes the load, it can interrupt the 24-hour baseline or move the measurement plane. Keep each unit powered and correctly terminated while switching measurement channels. Record power-off and restart times, temperature, sample number, counter reference and gate. Verify a new station or reference before comparing readings separated by only a few hundredths of a hertz.

Backup power creates another mismatch. Removing AC can blank a controller display while the oven still receives power. For retrace, confirm at the oscillator supply pin that power was actually removed and start the 24-hour interval at that point. A project concerned with battery-backed transfer can run a separate powered transfer test and record the supply minimum. Each report should identify the condition the oscillator actually experienced.

A weekend test can schedule the instrument to save readings at the agreed times and monitor reference availability. If a reading was taken 59 minutes after restart, retain that timestamp. Repeat if needed at the approved time. Agree with the supplier on any permitted time tolerance before adding it to the purchasing requirement.

Make the first-article test executable

One complete OX-176 retrace test takes 49 hours from first power-on. Put the sample arrival, chamber reservation and report delivery in the purchasing plan. Five units may run in parallel if each keeps the documented supply and termination. Retain both original readings, their difference, temperature, control voltage and timestamps for each unit. If combined uncertainty reaches 0.03 Hz against a 0.10 Hz limit, agree on the retest and disposition of boundary results beforehand. A lab unable to resolve the limit can request supplier test evidence under the agreed conditions.

State the 5 V or 12 V configuration, whether tuning is required and the sine-wave output. For a tunable part, record the real control setting and keep it the same at both readings. Retuning after restart can hide the condition the test is meant to observe. If the system must resume after 20 minutes, capture the second power cycle at 20 minutes too and evaluate it against a separate system limit. Approve the exact 10 MHz OCXO and board revision after both manufacturer retrace and the instrument's own recovery requirement have evidence.