Does the Microchip OX-224 10 MHz OCXO's −30 dBc harmonic limit satisfy a −25 dBm requirement?

A shared-reference instrument needs a replacement 10 MHz clock, and its receiving interface limits the second harmonic to −25 dBm. A supplier's OX-224 data sheet lists a maximum harmonic level of −30 dBc. Both numbers look similar and sometimes appear in the same quote field. For this 10 MHz oven-controlled oscillator, convert the carrier-relative limit to absolute power at the specified plane before deciding whether the part is suitable, needs a tighter guarantee or needs filtering. Preserve both units and the measuring point on the order.

Start with the largest allowed carrier

Microchip's RF output table specifies a sine wave into a nominal 50 Ω load. At 3.3 V and 50 Ω, its carrier limits are 5 dBm minimum, 7 dBm typical and 9 dBm maximum. The harmonic row gives a maximum of −30 dBc. A component at that relative limit would measure −25, −23 or −21 dBm for those three carrier levels, respectively. Add the relative dBc value to the carrier's dBm value. At the maximum carrier, the public harmonic numbers support a −21 dBm upper bound. That is 4 dB above the project's −25 dBm limit.

At a 9 dBm carrier, the second harmonic must be at or below −34 dBc to meet −25 dBm at that same reference plane. A sample with a 6 dBm carrier and −29 dBm second harmonic measures −35 dBc and passes in that setup. That result belongs to the sample and its load. A different shipment could still fall within the published limits without meeting the interface requirement. Put the exact harmonic requirement and test conditions in the purchase attachment if they are to govern lot release.

Give spurious signals their own line. The data sheet lists −80 dBc for spurious output, equivalent to a −71 dBm upper bound if the carrier is 9 dBm. For a lower system limit, define the frequency range, resolution bandwidth and reference plane. The second harmonic is at 20 MHz and the third at 30 MHz, while a power converter or another reference path could introduce other narrow lines. Combining “harmonics and spurious” under a single −80 dB value makes the inspection ambiguous.

Include the filter loss in the carrier budget

Suppose a candidate network has at least 6 dB attenuation at 20 MHz and no more than 1.2 dB insertion loss at 10 MHz. Assuming suitable load conditions and separate treatment of coupling effects, the −21 dBm second-harmonic bound becomes −27 dBm. That leaves 2 dB against the −25 dBm requirement. The minimum 5 dBm carrier can fall to 3.8 dBm, which is 0.2 dB below a hypothetical 4 dBm receiver minimum. Recalculate both limits before selecting that filter.

Identify the measurement plane: oscillator output, filter output or external interface. Cable loss, impedance and supply interference can differ. Record the correction at each measured frequency. If the harmonic budget has 2 dB of margin and measurement uncertainty is ±1.5 dB, only 0.5 dB remains after that allowance. Agree on the release and retest rule for a borderline reading before the supplier and receiving lab apply different loss corrections.

Establish whether the system limits apply during warm-up or only after settling. The manufacturer's general performance notes specify warm-up and typical supply, control, load and temperature unless noted. Record the start of timing, the measured supply and when RF output first appears. If two units share a supply or distribution board, system validation can also look for spurious output on one channel while the other starts. Keep the component record tied to the specified load and settled single-output condition.

For an outside lab, require the carrier measurement, harmonic difference relative to that measured carrier and absolute harmonic power. For example, −26 dBm at the harmonic with a 3 dBm carrier gives −29 dBc. It meets the example absolute limit but exceeds the public relative harmonic maximum by 1 dB, and the carrier is below the public minimum. Investigate port loading, connections, attenuation calibration and device supply before deciding which results apply.

Release samples under the actual project conditions

For the 10 MHz OCXO first article, confirm supply, load, warm-up and exact ordering code. Record carrier, second harmonic and spurious results separately. Set analyzer attenuation to avoid creating second-harmonic distortion in the input stage, while keeping the noise floor below the intended measurement. Repeat at a different attenuation with suitable reference-plane corrections and check whether the relative reading remains consistent. Use a separate control setup if required. Link the results to the part label, lot, filter state and instrument settings.

Approve the board configuration if the applicable manufacturer conditions and receiver budget cover both limits. Where the public data leave a 4 dB gap, request a confirmed tighter second-harmonic option or validate the complete filter route. Engineering samples can still be useful while those decisions remain open. The −30 dBc table entry alone does not support a production guarantee of −25 dBm.

References