An instrument board has a DAC for oscillator tuning. Its replacement proposal specifies a Microchip OX-228 10 MHz OCXO. The new specification mentions a nominal EFC voltage by the initial accuracy entry. Someone takes that as evidence of a tuning input and assigns the old control signal to one of the pads. The ordering information gives a fixed-frequency configuration and several pads are reserved for internal use. Identify where correction is supposed to occur so first article can distinguish a compliant oscillator from a system that can actually be calibrated.
Check the ordering field and pin drawing
Microchip lists standard CMOS frequencies of 5, 10, and 20 MHz and a standard sine output of 10 MHz. The frequency control code 0 means no tuning. The electrical table for each output describes a fixed OCXO with no adjustment. If the mention of nominal EFC voltage in the accuracy conditions causes doubt, ask the manufacturer to confirm the specific order code. A test-condition label does not establish an external adjustment pin. Do not infer its location from another series.
The package page identifies pin 3 as the supply, pin 4 as RF output, and pin 7 as case ground. Pin 1 is internally connected and must remain open. Pins 5 and 6 are also internally connected and have to remain open. Pin 2 is NC. If the old board connects any of these pads to ground, a probe, or a DAC, have engineering review the land pattern and nets before fitting the part. When electrical tuning is required, source a part with a defined tuning input, control voltage range, and pull capability. For a fixed reference, carry the no-tuning requirement and reserved pins into assembly and fixture instructions.
Calculate the fixed error before assigning correction to another stage
The public tables list ±0.2 ppm initial accuracy at shipment. At 10 MHz, this is ±2 Hz. If the next module must receive a physical reference within ±1 Hz and the board cannot adjust it, the specification leaves a 1 Hz shortfall. In an otherwise ideal four-times frequency multiplication, ±2 Hz at the input becomes ±8 Hz at 40 MHz. The fractional error remains 0.2 ppm. Add temperature, aging, and downstream errors under their applicable definitions. A tight temperature stability grade does not guarantee that uncorrected absolute frequency is within 1 Hz.
If the design can correct a calculated system value downstream, test with a known offset. A reading of 10,000,001.2 Hz is +0.12 ppm. Suppose the downstream correction has a 0.2 Hz effective step referred to the input. Ideal rounding leaves at most 0.1 Hz, before instrument errors, aging, or implementation effects. Keep both the corrected output and the unchanged physical oscillator reading. For another device connected directly to the same reference through a cable, the digital correction only affects the processed path. Specify a separate acceptance limit for the uncorrected physical output.
Verify output format and tuning capability at receipt and first article
The OX-228 output variants need different fixtures. The CMOS version uses a 15 pF load for logic levels. The sine version specifies 45 to 55 Ω and 0 to 6 dBm at 3.3 V. When measuring the sine format in 50 Ω, record the termination and measurement point. For a sinusoid, 0 dBm is 1 mW, about 224 mV RMS and 632 mV peak-to-peak in 50 Ω. This conversion does not justify loading the CMOS output with a 50 Ω instrument. If two lots of 10 MHz parts give different amplitudes, verify both the format and the fixture.
Verify each physical output position at the end of line. An instrument may have a reference connector for external equipment and a separate multiplied acquisition clock. Retain both uncorrected readings, the corrected result, and configuration readback. If the acquisition clock can be corrected to 0.1 Hz but the independent 10 MHz connector remains 1.2 Hz high against a ±1 Hz requirement, the instrument still misses one requirement. It needs a tunable upstream reference or a separately approved external specification. End-of-line calibration does not establish a screened component grade. That requires an agreed shipping criterion and a long-term limit. Repair instructions should distinguish the adjustable location and the test-only output so a technician does not apply voltage to a reserved pad.
Put the full manufacturer code, supply, waveform, operating temperature, and control option 0 on the order. Attach the applicable drawing and written clarification to the board file. Incoming inspection verifies identity and lot. First article checks the specified load, levels, and frequency. If the previous tuning trace is connected to a reserved pad, retain the original board and specimen condition. Do not use DAC sweeps to try to establish how it works. Reverify the redesigned board under its approved conditions. For substitution into an explicitly tunable position, send a new request specifying minimum pull, direction, and usable voltage range before buying volume.