How long after calibration can an OCXO go before it must be calibrated again? A single calibration report cannot establish the direction or rate of this unit's drift. Start with several timestamped comparisons against a traceable reference under the same warm-up, environmental and supply conditions.
Suppose field history shows approximately 0.01 ppm of drift per month over one observed interval. The system budget for calibration residual and aging together is 0.20 ppm. If 0.05 ppm must be reserved for post-calibration uncertainty, the linear model gives 0.15 ÷ 0.01 = 15 months. Assigning the entire 0.20 ppm to aging would give 20 months. The difference comes from how the budget is allocated.
When to check ahead of schedule
After transport, extended storage or a change in cooling, make a comparison using the usual warm-up conditions. Distinguish a restart step from sustained drift. Beyond the observed period, use the calculation only to schedule the next measurement. The manufacturer's guarantee period, the reference calibration validity and the spread of subsequent measurements all affect the formal recalibration date. Example Taitien OCXO aging specification