Changing the threshold on the same 200 MHz waveform can change the measured jitter. The instrument locates a threshold crossing on a noisy slope. A flatter slope gives the same voltage disturbance more effect on the crossing time.

Suppose the slope near the threshold is 0.5 V/ns, and the voltage moves by 10 mV. The local approximation gives Δt ≈ ΔV/(dV/dt) = 0.010 ÷ 0.5 ns = 20 ps. The same disturbance on a 1 V/ns segment gives an estimate of only 10 ps. Use the slope at the threshold. Averaging over the full edge can conceal shoulders or ringing.

One acquisition, two timestamp extractions

Save the original waveform and extract timestamps in software at two reasonable thresholds. This keeps the noise and samples consistent. Record the bandwidth, thresholds, local slopes and statistical definition of the time spread. If the 10 mV represents peak-to-peak noise, interpret the estimate on the same peak-to-peak basis. Before assessing oscillator RMS jitter, use a stable reference source to check the instrument's own timing spread. SiTime: clock jitter definitions and measurement