The same 10 MHz oscillator initially measures 10,000,002 Hz and later measures 10,000,005 Hz. It gained 3 Hz between measurements, or +0.3 ppm relative to nominal. The later reading is +5 Hz, or +0.5 ppm, above 10 MHz. Give these two quantities separate columns in the report.
When can you record the 3 Hz change as aging?
If the first measurement followed a full warm-up at 25 °C but the retest was taken straight out of the chamber, the 3 Hz may include thermal recovery. Restore the original temperature, supply, and load as prescribed, wait for thermal stability, and use the same reference method. Use the denominator specified by the manufacturer or inspection specification, and identify it in the table header.
For each sample, retain the ID, initial reading, final reading, elapsed time, and sign. If one device drifts up and another down, the lot mean may move very little. Paired readings for each sample reveal their individual behavior. Record power interruptions, transport, changes of test setup, and recalibration of the reference on the corresponding dates. Taitien: example OCXO aging limits and observation conditions