For a high-stability OCXO, drift in ppb per day and drift during the first year answer different questions. Multiplying a 24-hour guarantee that applies after operation has stabilized by 365 leaves out what happened after initial power-on.
One datasheet can specify two different observation periods
For each specification line, record the start of the observation period, its duration, the reference temperature and the supply conditions. The Taitien NF-10M-7000 series lists aging separately under frequency stability, where you can check the applicable time limits. For a different ordering option, use the period guaranteed for that part. Taitien NF-10M-7000 series frequency stability
If the application can go offline only once a year, request the maximum change over that interval and the applicable conditions. If the supplier provides only a daily figure for stabilized operation, keep the corresponding annual budget entry open.
Plot drift in the field for each unit. Repeat measurements with the same warm-up time, temperature and reference, and mark extended power outages, repairs and transport events. Use the trend to schedule recalibration. If you use it to predict drift, state the extrapolation interval and keep it separate from the manufacturer's guarantee.