An acceptance window of 50 MHz ±5 Hz gives a tolerance of only ±0.1 ppm. If the counter's reference uncertainty is already about 0.2 ppm, it contributes about 10 Hz to this measurement. Even a display that resolves 0.1 Hz cannot make borderline samples easy to assess on this reference.
Check the instrument's reference-source selection. Confirm that the external standard's frequency and amplitude meet the input specifications, and check that the instrument stays locked. Record the reference ID, calibration date, gate time, temperature, and other applicable conditions. A more accurate reference cannot improve the reading if the cable is connected but the instrument still uses its internal timebase.
When a reading is close to a limit, evaluate the timebase, triggering, and repeatability under the inspection decision rule before releasing the sample or transferring the test to a more suitable setup. Include these conditions when requesting the supplier's sample records so that a difference of a few hertz can be explained. Keysight: frequency counter accuracy and timebase