The oscilloscope may show only a 400 mV differential signal while each probe input is close to 2 V. Before probing a 5032 differential oscillator, select the probe range using the expected input voltages. Package size says nothing about common-mode voltage.
Suppose the two lines average 2.0 V at one instant and P is 0.4 V above N. P is then 2.2 V relative to ground, and N is 1.8 V. Compare both values with the operating range from each probe input to ground for the selected setting. Check the common-mode range, differential range and effective bandwidth separately. The 0.4 V is an instantaneous difference. To determine peak-to-peak voltage, you also need the difference in the other state.
If the clipping changes with the probe range
Keep the board loading, trigger point and bandwidth unchanged and repeat the measurement with a setting that covers the required ranges. If necessary, use a suitable single-ended probe to inspect the individual voltages to ground. If the differential clipping disappears with a different range, check the measurement chain before deciding whether to return the sample. If it remains at the receiver, check driver swing and termination. Save the probe model, attenuation setting and connection method so the measurement can be repeated on another instrument. Tektronix: linear dynamic range of active probes