80 MHz oscillator: how many cycles does the oscilloscope record cover?

If a fault occurs only once every few milliseconds, how long does one clean 80 MHz waveform actually cover? At an actual sample rate of 1 GS/s, a record of 10,000 points spans about 10 µs. An 80 MHz clock has a 12.5 ns period, so the entire record contains only about 800 cycles.

Choose between continuous capture and an anomaly trigger

To retain 5 ms of continuous history around the fault, the same sample rate requires about 5 million points. After widening the time span, read back the actual sampling rate and per-channel memory depth, particularly when acquiring multiple channels at once. If the instrument reduces its sampling rate, narrow pulses may disappear from the capture.

If you only need to detect a short pulse below the receiver's tolerance, measure the normal pulse width at the receiver threshold first. Set an abnormal pulse-width trigger and save pretrigger data. Segmented memory is useful for storing fragments around several events. Use another channel to align the power, OE, or reset signal. Report the actual number of captures and observation window. The dead time between triggers was outside that coverage.