A 26 MHz crystal reads 26,000,250 Hz at incoming inspection. The purchase order calls for an initial tolerance of ±10 ppm, so the upper limit is 26,000,260 Hz. The displayed reading is 10 Hz inside that limit. If the expanded measurement uncertainty is 20 Hz, should the inspector release this unit? Check the exact frequency specification and test conditions first, then apply the decision rule the buyer and supplier have agreed to use.
Convert ppm to the operating window used by the fixture
At 26 MHz, 1 ppm equals 26 Hz. The ±10 ppm tolerance gives ±260 Hz, or an interval from 25,999,740 to 26,000,260 Hz. Epson's TSX-3225 order code X1E000021014317 is an example. Section 2 of its individual specification lists a 26.000000 MHz fundamental part, a 10 pF load and an initial tolerance of ±10 ppm at +25 °C. Temperature stability from −20 to +75 °C and first-year aging have separate rows. Section 5 identifies suffix 17 as 4,000 pieces per reel. A quote that only says TSX-3225 and 26 MHz leaves the test team without the exact specification. Epson TSX-3225 individual specification, Sections 2 and 5.
Enter the fixture load, temperature, excitation condition, and test circuit beside the frequency limits. If the inspection method differs from the supplier's specified method, agree on how to correlate results before making a release decision. For example, a fixture with unverified load might indicate 25,999,750 Hz several times in a row and still carry a systematic offset. Repeated readings help estimate repeatability. Use a calibrated reference or a characterized comparison setup to check a persistent offset.
Decide how readings near the limit will be handled
Suppose the buyer and supplier use a three-way decision for this example. The expanded uncertainty is ±20 Hz. Accept when the measurement interval is entirely within tolerance, reject when it lies entirely outside, and review a result whose interval crosses a limit. Under this illustrative rule, the direct acceptance interval for displayed readings is 25,999,760 to 26,000,240 Hz. Use the actual bench evaluation to set the uncertainty. Other approved agreements may apply a different risk-sharing rule or accept by the displayed value, so document the chosen rule on the inspection plan.
| Fixture reading | Interval with ±20 Hz | Illustrative disposition |
|---|---|---|
| 26,000,235 Hz | 26,000,215 to 26,000,255 Hz | Accept |
| 26,000,250 Hz | 26,000,230 to 26,000,270 Hz | Review at boundary |
| 26,000,285 Hz | 26,000,265 to 26,000,305 Hz | Handle under the agreed nonconformance procedure |
An assumed 0.2 ppm time-base specification translates to 5.2 Hz at 26 MHz. That is one contribution under its stated conditions. The counter's function, gate, input signal, external reference lock and warm-up can affect the result. Fixture loading also needs attention. A display that shows single hertz does not establish single-hertz accuracy. For a result near a limit, repeat with a verified fixture using the agreed method, and keep the raw readings, reference setup and decision in the record. Keysight frequency measurement uncertainty and time-base specifications · ILAC G8 guidance on decision rules.
When the reference reads high only on the product board
Suppose the same crystal measures 26,000,130 Hz in the verified fixture and 26,000,300 Hz when inferred from the product board's buffered output. Those values are +5 ppm and about +11.54 ppm. The 170 Hz difference prompts a check of the board load, internal capacitor settings, clock path and measurement reference. Judge the fixture result against its own receiving interval. If the fixture also exceeds the limit, verify load, temperature, drive and exact order code, and retain the other sampled units' results. Disposition of one unit follows the agreed decision rule. Lot disposition follows the project's sampling plan. A useful handoff gives engineering and purchasing the part number, lot, fixture and board readings, corresponding limits, and who is responsible for the next check.
Retain the measured value and the disposition as two fields. That keeps a result near the limit available for later review if the method changes or if the supplier measures the same unit using a different fixture. Assign the review to receiving quality until the evidence points to a board issue.