Two RTC one-second pulses overlapping on an oscilloscope show that their phases are close at that instant. Use longer persistence or record consecutive timestamps to see whether the clocks run at the same rate. A clock continuously running 10 ppm fast ideally gains 36 ms in 1 hour, or about 0.864 s in one day.
A slope and a step
Keep the reference fixed and define the positive direction for phase difference. If the convention is elapsed time on the device under test minus elapsed time on the reference, a positive slope means the device runs fast. Subtracting edge arrival times can produce the opposite sign because the faster clock's edge arrives earlier. For long observations, number the one-second edges so a phase difference exceeding one second does not cause adjacent pulses to be paired incorrectly.
An approximately linear slope can estimate the average frequency error. Periodic steps may come from digital correction or active synchronization, so record the configuration and step times. Compare windows of the same duration at similar temperatures before and after calibration, and cover a complete correction cycle. Calendar timekeeping and smooth one-second edge timing are separate acceptance items. Report each according to the application. Epson RX8900SA manual: temperature compensation and output
