A counter reads 40,000,600 Hz from the buffered clock on a 40 MHz board. Engineering is ready to fit larger capacitors, while purchasing suspects the latest crystal lot. Start by converting the board reading to +15 ppm and try one reversible load change. That one reading includes the crystal, the load presented by the board, and the measuring system.
Pick a point that represents the uncorrected reference
Trace the clock tree and read back its configuration. A verified divide-by-four clock at 10,000,150 Hz corresponds to 40,000,600 Hz at the reference. If the point is after a PLL, an automatic correction stage, or a source multiplexer, establish what it actually measures. Use a supported output driven by the crystal reference if possible. Read back the internal capacitor code too. Two boards with the same external parts can have different capacitance settings in firmware. Let the board and counter reference stabilize. An assumed 2 ppm time-base error represents 80 Hz at 40 MHz. A longer counter gate can improve resolution, while reference accuracy still needs its own check.
Make one load change and reverse it
The example below uses the same crystal, board, connection and instrument. Change either a pair of external capacitors or a documented internal trim step, one at a time, and restore the initial state afterward. In this hypothetical case the effective load increases by approximately 0.50 pF. Allow the board to settle after rework and keep supply and test settings consistent.
| State | Frequency referred to the crystal | Error from 40 MHz |
|---|---|---|
| Original | 40,000,600 Hz | +15.00 ppm |
| About 0.50 pF extra load | 40,000,280 Hz | +7.00 ppm |
| Original restored | 40,000,590 Hz | +14.75 ppm |
The load change moved the reference down about 320 Hz, and the restored reading is 10 Hz from the original. If the measurement reliably resolves the larger shift, the experiment supports tuning the load. A 320 Hz difference at 40 MHz is 8 ppm. Dividing by the estimated 0.50 pF gives about −16 ppm per pF over this small interval. Removing all 600 Hz would take about 0.94 pF if that slope stayed constant. Use this only to select the next settings to try. A trim register may not have uniform steps, the crystal response is not linear over a wide range, and any setting must also pass startup tests. If the restored reading continues to drift, check temperature and the counter reference. If a higher load increases the displayed frequency, trace the measurement point and clock correction settings before adjusting again.
Two 40 MHz parts from one series can have different temperature limits
Epson's FA-128 specifications provide a concrete purchasing example. The fundamental 40 MHz, 2.0 × 1.6 mm Q22FA1280010118 specifies a 10 pF load. Q22FA1280033218 specifies 12 pF. Both list an initial tolerance of ±10 ppm at +25 °C. The 10 pF part's temperature stability is ±20 ppm from −40 to +85 °C. The 12 pF part's separate sheet specifies ±10 ppm from −20 to +70 °C. A quote giving only FA-128, 40 MHz and 10 ppm leaves out the difference. Compare each exact order code and its first specification page against the design conditions, and check availability separately. Epson 10 pF individual specification · Epson 12 pF individual specification.
For a board-only change, preserve the original and candidate capacitor values, firmware and measured frequency. Verify startup at the required supply and temperature limits. For a crystal change, verify the new part's rated conditions first, then check whether the current board can meet frequency and startup targets across the intended lots. A centered room-temperature sample is a good starting point for that qualification. Epson's load capacitance and pulling explanation · Keysight frequency-counter uncertainty and time base.