If two counters disagree by a dozen or so hertz, leave the oscillator adjustment alone first. Feed the same signal to A and B simultaneously through a suitable splitter. Confirm that the amplitude, termination, trigger threshold, counting mode, and gate are comparable. If using a tee, calculate the resulting parallel load.
| Change one condition at a time | What to observe |
|---|---|
| Switch both counters to the same verified external reference | Confirm that both counters are locked. If the difference shrinks substantially, check the original individual timebases first. |
| Keep the reference unchanged and swap the complete distribution branches | If the higher reading moves to the other instrument, check the splitter ports, cables, amplitude, and triggering. |
| The difference still follows the same instrument | Repeat with matching settings and check input channel, warm-up, and calibration status. |
Record the sign of A−B each time. It is easier to see whether the discrepancy follows the wiring when the sign is known than when the record says only that the readings differ by a dozen or so hertz. Even if A and B agree on a shared reference, both can carry the same absolute error. Check the external standard's calibration and uncertainty separately before releasing the product. Keysight: frequency counter measurement conditions