Oscillator frequency-error test: why do two instruments disagree by a dozen hertz or more?

If two counters disagree by a dozen or so hertz, leave the oscillator adjustment alone first. Feed the same signal to A and B simultaneously through a suitable splitter. Confirm that the amplitude, termination, trigger threshold, counting mode, and gate are comparable. If using a tee, calculate the resulting parallel load.

Change one condition at a timeWhat to observe
Switch both counters to the same verified external referenceConfirm that both counters are locked. If the difference shrinks substantially, check the original individual timebases first.
Keep the reference unchanged and swap the complete distribution branchesIf the higher reading moves to the other instrument, check the splitter ports, cables, amplitude, and triggering.
The difference still follows the same instrumentRepeat with matching settings and check input channel, warm-up, and calibration status.

Record the sign of A−B each time. It is easier to see whether the discrepancy follows the wiring when the sign is known than when the record says only that the readings differ by a dozen or so hertz. Even if A and B agree on a shared reference, both can carry the same absolute error. Check the external standard's calibration and uncertainty separately before releasing the product. Keysight: frequency counter measurement conditions