Specialist Supplier of Crystals & Frequency Control Componentssales@mecxtal.com.cn0755-26500570
MERLIN home / Rednote / Measurement and Acceptance

Comparing two low-noise OCXOs: how to tell when the phase-noise analyzer has reached its floor

Two low-noise OCXOs can produce nearly identical flat traces far from the carrier even when their actual noise differs. If measurement-channel noise approaches that of the device under test, it can hide the difference. A two-channel cross-correlation analyzer retains components common to both inputs while suppressing uncorrelated channel noise, helping resolve lower-noise sources.

Use the number of correlations as a diagnostic variable

Rohde & Schwarz gives an ideal sensitivity improvement of about 5 log₁₀(N) dB, where N is the number of correlations. Increasing N from one to one hundred improves sensitivity to uncorrelated instrument noise by about 10 dB in theory. This is a sensitivity improvement and cannot simply be subtracted from the measured device result. More correlations also take longer. Keep supply, carrier level, and environment stable throughout the run.

Keep the offset range, input level, and all other analyzer settings fixed. Save a trace at a low correlation count, then increase it step by step. If the background moves down, the original measurement was at least partly instrument-limited. If a discrete line stays at the same offset and amplitude, check the device and any common path. A shared reference, supply, or crosstalk can add correlated components to both channels.

Once the trace converges, check measurement headroom with a known lower-noise reference source or the manufacturer's recommended floor verification. If you cannot prove the floor is below the project limit, report the upper bound the setup can resolve, together with the correlation count and input conditions. Two traces that sit on the floor are insufficient to rank the devices. When comparing vendor data, check that the offset ranges, smoothing, and treatment of spurs match.Rohde & Schwarz: cross-correlation phase-noise measurement principles