A warmed reference runs at slightly different frequencies when horizontal and vertical. Stress direction can change quartz resonance, so manufacturers commonly specify fractional shift per unit of gravitational acceleration. Use a known gravity projection along the device axis and keep supply, cabling and thermal conditions fixed to separate this effect from drift.
Pair opposite orientations
For one axis, compare orientations parallel and antiparallel to gravity. The acceleration component changes by about 2 g. If a 10 MHz source shifts by 0.04 Hz between settled positions, that is 4 ppb, or about 2 ppb/g for that axis. A horizontal-to-vertical move can change projections on two axes, so dividing the raw result by two does not give a single-axis coefficient.
After each flip, let mechanical and thermal conditions settle. Avoid cable strain and airflow changes. Measure positive, negative and positive again to check continuing drift. Repeat for the other two perpendicular axes and retain orientation, mounting and signed differences. A static flip measures sensitivity to a gravity reversal and cannot establish shaker response at every frequency.
Calibrate in the actual mounting orientation for a fixed installation. For equipment that moves, include the orientation range in its budget. In the RFQ, specify a single-axis maximum or a three-axis metric, along with frequency and conditions. If the fixture cannot produce an accurate axis reversal, measure the angle and correct using the gravity projection.Microchip crystal acceleration sensitivity measurement method