Two 10 MHz oven oscillators are compared overnight. At short averaging times their Allan deviations look similar, but one curve turns upward at long times. Plot the original fractional frequency against time. A sustained frequency slope enters the Allan deviation. Keeping the raw statistic, the fitted drift rate, and the Hadamard deviation together helps distinguish variability from actual accumulated frequency movement.

How a small slope appears on the plot

Take continuous fractional frequency y(t) = a + Dt. Averages in adjacent windows of length τ differ by Dτ. Linear frequency drift alone contributes an Allan deviation of |D|τ/√2. With the illustrative slope D = 3 × 10⁻¹¹/s, the deviation is about 2.12 × 10⁻¹⁰ at τ = 10 s and 2.12 × 10⁻⁸ at τ = 1000 s. The latter corresponds to about 0.212 Hz when scaled to 10 MHz. It represents a normalized difference between window averages. The same slope changes the actual frequency by about 1.08 Hz in one hour, an important effect that a short-time statistic alone would miss.

Hadamard deviation uses second differences of three successive frequency averages. Subtract twice the middle average from the sum of the first and third. For averages of 0.0, 0.3, and 0.6 ppb, the result is zero. If the last value is 0.8 ppb, the difference is 0.2 ppb. For this one illustrative three-window term, the normalized deviation is 0.2/√6, about 0.082 ppb. A real curve requires more data. The longest-τ point can have few independent degrees of freedom.

Check whether the bench produced the slope

Compare the frequency timeline with power-up, chamber actions, supply, and reference status. Archive the warm-up segment even if the test calls for statistics calculated from a later, predefined start time. Heating, HVAC cycles, and aging can all resemble a line over limited intervals. Plot slopes for contiguous equal-length portions to see whether direction and magnitude persist. If a second verified reference gives a different slope, or multiple devices move together, investigate the reference and shared environment. A relative-frequency record directly measures only the difference between the two sources.

Use contiguous, evenly spaced data for standard estimates. Missed readings, unlocks, and range changes must not become fake adjacent samples. Keep step events annotated and handle them under the test procedure. A prespecified linear fit over a defined interval can be used to calculate residual statistics alongside the Hadamard deviation. Steps, nonlinear heating, and cycling can affect the latter even though linear drift is rejected.

Which result supports acceptance?

If the specification calls for overlapping Allan deviation at τ = 1, 10, and 100 s, use that statistic under the specified temperature, bandwidth, and warm-up conditions. Hadamard results can explain a long-τ trend. A system that depends on the frequency excursion two hours after start also needs the original trace and the observed excursion. At the example slope, the two-hour movement at 10 MHz is 2.16 Hz. If the application tolerates 1 Hz, the drift requirement needs evaluation even when the residual curve is excellent.

Keep the raw trace, statistic, τ values, overlap setting, fit interval, slope, temperature, and reference conditions. Include record length and available confidence information at long τ so another engineer can reproduce the calculation and check the actual frequency budget.

Align frequency units before sharing calculations across benches. A counter may export absolute hertz, a script may export a dimensionless difference from nominal, and another report may already scale the result to ppb. With a fixed nominal frequency, subtract it from the measured frequency, then divide by it. The number of displayed decimal places does not document this conversion. Spot-check selected points manually from averages over actual τ-length windows and their differences. This catches confusion between seconds and sample counts.

For batch testing, save the trace of a check unit measured against the same reference. If the check unit also trends upward, stabilize the thermal environment or repeat against a different verified reference. A rise limited to one serial number with no corresponding supply event justifies extending that sample’s run and examining another power cycle. Plot the residuals before deciding whether a linear aging interpretation fits.